Buddhavarapu Athreya
GenesisCare, Epping, Victoria, Australia.
J Appl Clin Med Phys. 2024 Mar;25(3):e14203. doi: 10.1002/acm2.14203. Epub 2023 Nov 8.
The aim of this study was to investigate the suitability of three radiochromic film analysis software for stereotactic radiotherapy patient-specific quality assurance (PSQA): FilmQA Pro v5.0, SNC Patient v6.2, and eFilmQA v5.0.
Film calibration was conducted for each software followed by three sets of measurements. The first set assessed calibration accuracy by comparing measured and delivered doses at increments different from those used for calibration. The second set used each software to conduct PSQA through gamma analysis on 10 stereotactic radiosurgery (SRS) and stereotactic body radiation therapy (SBRT) patients. The third set utilized SNC Patient and eFilmQA to carry out gamma analysis on a collection of four digital test images, eliminating delivery and scanning uncertainties from impacting the analysis. Key supporting features within each software for conducting gamma analysis were identified.
Overall, FilmQA Pro and eFilmQA were deemed comparable and favoured over SNC Patient due to the presence of key features such as triple-channel dosimetry, auto-optimization, and dose scaling. FilmQA Pro has a substantial user base and established reputation. eFilmQA, having been introduced more recently, serves as a viable alternative to FilmQA Pro, having been further refined for stereotactic radiotherapy PSQA.
This study investigated the suitability of three film analysis software (FilmQA Pro, eFilmQA, and SNC Patient) for stereotactic radiotherapy PSQA. Results from the investigation indicated that both FilmQA Pro and eFilmQA are comparably suitable and are preferred over SNC Patient. Both FilmQA Pro and eFilmQA are recommended for radiotherapy clinics.
本研究旨在调查三种放射变色胶片分析软件用于立体定向放射治疗患者特异性质量保证(PSQA)的适用性:FilmQA Pro v5.0、SNC Patient v6.2和eFilmQA v5.0。
对每个软件进行胶片校准,然后进行三组测量。第一组通过比较在与校准所用增量不同的增量下测量剂量和输送剂量来评估校准准确性。第二组使用每个软件通过对10例立体定向放射外科(SRS)和立体定向体部放射治疗(SBRT)患者进行伽马分析来进行PSQA。第三组利用SNC Patient和eFilmQA对一组四张数字测试图像进行伽马分析,消除输送和扫描不确定性对分析的影响。确定了每个软件中进行伽马分析的关键支持功能。
总体而言,FilmQA Pro和eFilmQA被认为具有可比性,并且由于具有三通道剂量测定、自动优化和剂量缩放等关键功能而优于SNC Patient。FilmQA Pro拥有大量用户群体和既定声誉。eFilmQA是最近推出的,作为FilmQA Pro的可行替代方案,已针对立体定向放射治疗PSQA进行了进一步优化。
本研究调查了三种胶片分析软件(FilmQA Pro、eFilmQA和SNC Patient)用于立体定向放射治疗PSQA的适用性。调查结果表明,FilmQA Pro和eFilmQA都同样适用,并且优于SNC Patient。建议放射治疗诊所同时使用FilmQA Pro和eFilmQA。