Abstract

Factors detracting from the reliability of faecal egg counts based on the McMaster technique include variation in flotation time (interval between loading chamber and counting eggs) and sample dilution (ratio of faecal material to salt solution). We recommend standardization of both these variables as normal laboratory procedure, and propose optima of a 30 minute flotation time and a sample dilution of 15 ml salt solution/g faeces for use of the McMaster technique in the estimation of the fecundity of Heligmosomoides polygyrus (Nematoda).