Boone J M
Med Phys. 1986 Nov-Dec;13(6):861-8. doi: 10.1118/1.595811.
The concept of the equivalent spectrum (Seq) is introduced, where Seq is defined as an idealized energy spectrum which results in identical attenuation properties as the actual spectrum of a given x-ray tube. Seq is described by two parameters, the equivalent aluminum filtration Aleq and the equivalent kilovoltage kVeq, which are evaluated iteratively using aluminum attenuation data. Published spectral data are used to demonstrate the merits of the equivalent spectrum technique, and experimental results are also presented. The technique is shown to be sensitive to changes in total filtration and, indirectly, to the waveform properties of the x-ray system. The use of the Seq may result in a better description of spectral properties for reporting purposes, and aid in standardizing radiographic techniques within an institution.
引入了等效谱(Seq)的概念,其中Seq被定义为一种理想化能谱,它产生的衰减特性与给定X射线管的实际能谱相同。Seq由两个参数描述,即等效铝过滤值Aleq和等效千伏值kVeq,它们使用铝衰减数据进行迭代评估。已发表的能谱数据用于证明等效谱技术的优点,同时也给出了实验结果。结果表明,该技术对总过滤的变化敏感,并且间接对X射线系统的波形特性敏感。使用Seq可能会为报告目的更好地描述能谱特性,并有助于在机构内规范放射技术。