Grove D E, Stein R A, Nosil J
Med Phys. 1986 Nov-Dec;13(6):949-53. doi: 10.1118/1.595824.
The dependence of the left ventricular ejection fraction calculation on the background correction in gated cardiac blood pool studies is examined by analyzing a simple cardiac model. It is shown that the ejection fraction is relatively insensitive to the background correction when an accurately determined variable region of interest is used, but that with a fixed region of interest the background correction must be accurately estimated. A similar ejection fraction-background correction relationship is obtained in the analysis of clinical cardiac data. Based on the cardiac model, two methods are suggested which, with further investigation and refinement, may lead to a more accurate and consistent technique for determining ejection fraction and background.
通过分析一个简单的心脏模型,研究了门控心血池研究中左心室射血分数计算对背景校正的依赖性。结果表明,当使用准确确定的可变感兴趣区时,射血分数对背景校正相对不敏感,但使用固定感兴趣区时,必须准确估计背景校正。在临床心脏数据的分析中也得到了类似的射血分数-背景校正关系。基于该心脏模型,提出了两种方法,经过进一步研究和完善,可能会产生一种更准确、更一致的测定射血分数和背景的技术。