视频眼动描记系统与交替棱镜遮盖试验测量斜视的对比分析。

Comparative Analysis of Strabismus Measurement Using a Video Oculagraphy System and Alternate Prism Cover Test.

机构信息

Departments of Optometry.

Ophthalmology, Vissum (Miranza Group), Alicante, Spain.

出版信息

Asia Pac J Ophthalmol (Phila). 2023;12(6):582-590. doi: 10.1097/APO.0000000000000641. Epub 2023 Nov 17.

Abstract

PURPOSE

The aim of the study was to compare distance vision measurements obtained with video oculography (VOG) and an alternate prism cover test (APCT).

DESIGN

Single-center, retrospective, and cross-sectional.

METHODS

Eighty-seven subjects with strabismus were included. All patients underwent an optometric examination. The measurement of strabismus in distance vision was performed with the best optical correction using the APCT and the VOG GazeLab. Subjects were divided according to the type of strabismus; 41 were esotropic and 46 were exotropic.

RESULTS

The general comparison of APCT and VOG showed a mean difference of 0.60±2.74 prism diopters (Δ), not observing statistical differences between both methods ( P =0.059) and presenting a correlation of 0.976 ( P <0.001). Using APCT, the mean amount of esotropia for the group was 18.31±11.64 Δ and that of exotropia was 19.62±8.80 Δ. Compared with the VOG, the mean value of esotropia for the group was 18.65±11.65 Δ and that of exotropia was 19.48±8.62 Δ. The means are statistically different for the esotropia group ( P =0.005) but not for the exotropia group ( P =0.318). There was a high direct correlation between the methods of measurement for both the esotropic ( R =0.980; P <0.001) and exotropic patients ( R =0.975; P <0.001). Bland-Altman analysis shows a mean difference of 1.37±2.76 Δ for the esotropia patients and 0.31±2.66 Δ for the exotropic patients, respectively.

CONCLUSIONS

This study demonstrated comparable results in measuring strabismus between VOG and APTC for esotropia and exotropia, with an excellent correlation between both methods and good agreement, especially in subjects with exotropia.

摘要

目的

本研究旨在比较视频眼动描记术(VOG)和交替棱镜遮盖试验(APCT)获得的远距离视力测量值。

设计

单中心、回顾性、横断面研究。

方法

纳入 87 例斜视患者。所有患者均接受眼科检查。使用 APCT 和 VOG GazeLab 进行最佳光学矫正后,测量远距离斜视。根据斜视类型将患者分为 41 例内斜视和 46 例外斜视。

结果

APCT 和 VOG 的总体比较显示,平均差异为 0.60±2.74 棱镜度(Δ),两种方法之间无统计学差异(P=0.059),相关性为 0.976(P<0.001)。使用 APCT,该组内斜视的平均斜视量为 18.31±11.64Δ,外斜视的平均斜视量为 19.62±8.80Δ。与 VOG 相比,该组内斜视的平均斜视量为 18.65±11.65Δ,外斜视的平均斜视量为 19.48±8.62Δ。内斜视组的平均值存在统计学差异(P=0.005),但外斜视组无统计学差异(P=0.318)。两种斜视患者的测量方法之间均存在高度正相关(内斜视:R=0.980;P<0.001;外斜视:R=0.975;P<0.001)。Bland-Altman 分析显示,内斜视患者的平均差异为 1.37±2.76Δ,外斜视患者的平均差异为 0.31±2.66Δ。

结论

本研究表明,VOG 和 APTC 对内斜视和外斜视的斜视测量结果具有可比性,两种方法之间具有极好的相关性和良好的一致性,尤其是在外斜视患者中。