Department of pharmacy, Hebei General Hospital.
Cardiovascular medicine, North China University of Science and Technology, Tangshan.
Blood Press Monit. 2023 Oct 1;28(5):284-288. doi: 10.1097/MBP.0000000000000662. Epub 2023 Jul 12.
This study aimed to validate the accuracy of DBP-6279B, an automated inflationary oscillometric upper-arm blood pressure (BP) monitor, in the sitting position according to the AAMI/ESH/ISO (81060-2 : 2018 + Amd.1 : 2020) universal standard protocol. SBPs and DBPs were measured simultaneously on the same arm in 88 adults (female : male = 47 : 41) with a mean age of 56.85 years using a mercury sphygmomanometer (two observers) and a DBP-6279B device (one supervisor). The AAMI/ESH/ISO 81060-2 : 2018 and Amd.1 : 2020 universal standards for the validation of BP-measuring devices in adults and adolescents were followed. A total of 259 valid pairs of data were used in the analysis. According to Criterion 1, the mean difference of SBP between the test device (DBP-6279B) and the reference device (the mercury sphygmomanometer) was 0.75 mmHg, with a SD of 7.66 mmHg. The mean difference in DBP was 1.13 mmHg, with a SD of 6.14 mmHg. The mean difference of both SBP and DBP was less than 5 mmHg, and the SD was less than 8 mmHg, which met the requirements. According to Criterion 2, the mean difference of SBP between the test device and the reference device was 0.85 mmHg, and the SD was 6.56 mmHg, which was less than 6.88 mmHg and met the requirements. The mean difference in DBP was 1.27 mmHg, and the SD was 5.42 mmHg, which was less than 6.82 mmHg and met the requirements. DBP-6279B fulfilled the requirements of the AAMI/ESH/ISO universal standard (ISO 81060-2 : 2018 + Amd.1 : 2020); hence, it can be recommended for both clinical and self/home BP measurement in adults and adolescents.
本研究旨在根据 AAMI/ESH/ISO(81060-2:2018+Amd.1:2020)通用标准协议,验证 DBP-6279B 这款自动充气式振动脉冲上臂血压(BP)监测仪在坐姿下的准确性。在 88 名成年人(女性:男性=47:41)中,使用汞柱血压计(两名观察者)和 DBP-6279B 设备(一名主管)同时在同一手臂上测量收缩压(SBP)和舒张压(DBP),平均年龄为 56.85 岁。遵循 AAMI/ESH/ISO 81060-2:2018 和 Amd.1:2020 成人和青少年血压测量设备验证的通用标准。分析中使用了总共 259 对有效数据。根据标准 1,测试设备(DBP-6279B)与参考设备(汞柱血压计)之间 SBP 的平均差值为 0.75mmHg,标准差为 7.66mmHg。DBP 的平均差值为 1.13mmHg,标准差为 6.14mmHg。SBP 和 DBP 的平均差值均小于 5mmHg,标准差均小于 8mmHg,符合要求。根据标准 2,测试设备与参考设备之间 SBP 的平均差值为 0.85mmHg,标准差为 6.56mmHg,小于 6.88mmHg,符合要求。DBP 的平均差值为 1.27mmHg,标准差为 5.42mmHg,小于 6.82mmHg,符合要求。DBP-6279B 符合 AAMI/ESH/ISO 通用标准(ISO 81060-2:2018+Amd.1:2020)的要求;因此,它可以推荐用于成人和青少年的临床和自我/家庭血压测量。