Acad Radiol. 2024 May;31(5):2118-2127. doi: 10.1016/j.acra.2023.10.016. Epub 2023 Nov 25.
Dual-energy subtraction (DES) imaging constitutes a technique that has demonstrated efficacy in enhancing the detectability of pulmonary nodules on chest radiographs. However, a simple and quantitative methodology for evaluating the clinical image quality of DES images is currently lacking. The objective of this study was to investigate the applicability of contrast-detail (C-D) phantom analysis to the visual clinical image quality evaluation of chest DES images.
We employed a custom-made phantom incorporating the C-D phantom and a multipurpose anthropomorphic adult chest phantom. Two phantom sizes were utilized to simulate standard- and large-bodied adult patients for each phantom. The custom-made phantom images were scored automatically using dedicated software, yielding an inverse image quality figure (IQF) value. The multipurpose anthropomorphic adult chest phantom was employed in a visual grading analysis (VGA) study that was conducted by two experienced radiologists and five radiological technologists. Each nodule placed in the chest phantom image was rated on a 4-point Likert scale.
A statistically significant correlation was observed between the VGA scores of the seven observers and the obtained IQF values.
The findings of this study suggest that DES image analysis of the C-D phantom possesses the potential to be utilized for the evaluation of clinical DES image quality based on chest lesion detectability.
双能量减影(DES)成像是一种已被证明能提高胸部 X 线摄影中肺结节检测能力的技术。然而,目前缺乏一种简单而定量的方法来评估 DES 图像的临床图像质量。本研究的目的是探讨对比细节(C-D)体模分析在胸部 DES 图像视觉临床图像质量评估中的适用性。
我们使用了一种定制的体模,其中包括 C-D 体模和多功能成人胸部体模。为每个体模模拟了标准体型和大体型成年患者的两种体模尺寸。定制体模图像使用专用软件自动评分,得出逆图像质量因子(IQF)值。多功能成人胸部体模用于由两名有经验的放射科医生和五名放射技术人员进行的视觉分级分析(VGA)研究。在胸部体模图像中放置的每个结节均按 4 分李克特量表进行评分。
七个观察者的 VGA 评分与获得的 IQF 值之间存在显著的统计学相关性。
本研究的结果表明,基于胸部病变检测能力,C-D 体模的 DES 图像分析有可能用于评估临床 DES 图像质量。