Radiology. 2012 Jan;262(1):298-304. doi: 10.1148/radiol.11110447. Epub 2011 Nov 4.
To determine the correlation between the clinical and physical image quality of chest images by using cadavers embalmed with the Thiel technique and a contrast-detail phantom.
The use of human cadavers fulfilled the requirements of the institutional ethics committee. Clinical image quality was assessed by using three human cadavers embalmed with the Thiel technique, which results in excellent preservation of the flexibility and plasticity of organs and tissues. As a result, lungs can be inflated during image acquisition to simulate the pulmonary anatomy seen on a chest radiograph. Both contrast-detail phantom images and chest images of the Thiel-embalmed bodies were acquired with an amorphous silicon flat-panel detector. Tube voltage (70, 81, 90, 100, 113, 125 kVp), copper filtration (0.1, 0.2, 0.3 mm Cu), and exposure settings (200, 280, 400, 560, 800 speed class) were altered to simulate different quality levels. Four experienced radiologists assessed the image quality by using a visual grading analysis (VGA) technique based on European Quality Criteria for Chest Radiology. The phantom images were scored manually and automatically with use of dedicated software, both resulting in an inverse image quality figure (IQF). Spearman rank correlations between inverse IQFs and VGA scores were calculated.
A statistically significant correlation (r = 0.80, P < .01) was observed between the VGA scores and the manually obtained inverse IQFs. Comparison of the VGA scores and the automated evaluated phantom images showed an even better correlation (r = 0.92, P < .001).
The results support the value of contrast-detail phantom analysis for evaluating clinical image quality in chest radiography.
通过使用经蒂尔技术防腐的尸体和对比细节体模,确定胸部图像的临床和物理图像质量之间的相关性。
使用人体尸体符合机构伦理委员会的要求。通过使用经蒂尔技术防腐的三具人体尸体评估临床图像质量,这导致器官和组织的柔韧性和可塑性得到极好的保存。结果,在图像采集期间可以使肺部充气,以模拟胸部 X 线摄影上所见的肺部解剖结构。使用非晶硅平板探测器获取对比细节体模图像和经蒂尔技术防腐的尸体的胸部图像。改变管电压(70、81、90、100、113、125 kVp)、铜过滤(0.1、0.2、0.3 mm Cu)和曝光设置(200、280、400、560、800 速度级)以模拟不同的质量水平。四名有经验的放射科医生使用基于欧洲胸部放射学质量标准的视觉分级分析(VGA)技术评估图像质量。使用专用软件手动和自动对体模图像进行评分,两者均产生逆图像质量因子(IQF)。计算逆 IQF 与 VGA 评分之间的 Spearman 秩相关系数。
VGA 评分与手动获得的逆 IQF 之间观察到统计学上显著的相关性(r = 0.80,P <.01)。比较 VGA 评分和自动评估的体模图像显示出更好的相关性(r = 0.92,P <.001)。
结果支持使用对比细节体模分析来评估胸部 X 射线摄影中的临床图像质量。