Suppr超能文献

放射科多个X射线摄影系统的胸部和骨盆X射线成像协议的标准化与优化

Standardisation and Optimisation of Chest and Pelvis X-Ray Imaging Protocols Across Multiple Radiography Systems in a Radiology Department.

作者信息

Abdi Ahmed Jibril, Jensen Kasper Rørdam, Pietersen Pia Iben, Jensen Janni, Hovgaard Rune Lau, Holmboe Ask Kristian Aas, Gregersen Sofie

机构信息

Department of Clinical Engineering, Region of Southern Denmark, Odense University Hospital, J. B. Winsløws Vej 4, 5000 Odense, Denmark.

Research and Innovation Unit of Radiology, University of Southern Denmark, Kløvervænget 10, 5000 Odense, Denmark.

出版信息

Diagnostics (Basel). 2025 Jun 6;15(12):1450. doi: 10.3390/diagnostics15121450.

Abstract

X-ray imaging protocols in radiology departments often exhibit variability in exposure parameters and geometric setups, leading to inconsistencies in image quality and potential variations in patient dose. : This study aimed to harmonise and optimise chest and pelvis X-ray imaging protocols by standardising exposure parameters and geometric setups across departmental systems, minimising radiation dose while ensuring adequate image quality for accurate diagnosis. : The image quality of five pelvic and three chest protocols across different radiographic systems was evaluated both quantitatively and visually. Visual image quality for both chest and pelvis protocols was assessed by radiologists and radiographers using the Visual Grading Analysis (VGA) method. Additionally, the quantitative image quality figure inverse () metric for all protocols was determined using the CDRAD image quality phantom. Moreover, the patient radiation dose for both chest and pelvis protocols was evaluated using dose area product (DAP) values measured by the systems' built-in DAP metres. : Different quantitative image quality and radiation dose to patients were achieved in various protocol settings for both chest and pelvis examinations, but the visual image quality assessment showed satisfactory image quality for all observers in both the pelvis and chest protocols. The selected protocols for harmonising chest radiography across all imaging systems result in reduced radiation exposure for patients while maintaining adequate image quality compared to the previously used system-specific protocol. : The clinical protocol for chest and pelvis radiography has been standardised and optimised in accordance with patient radiation exposure and image quality. This approach aligns with the ALARA (As Low As Reasonably Achievable) principle, ensuring optimal diagnostic information while minimising the radiation risks.

摘要

放射科的X射线成像协议在曝光参数和几何设置方面常常存在差异,导致图像质量不一致以及患者剂量的潜在变化。本研究旨在通过跨部门系统标准化曝光参数和几何设置来协调和优化胸部及骨盆X射线成像协议,在确保获得足够图像质量以进行准确诊断的同时,将辐射剂量降至最低。对不同射线照相系统的五个骨盆和三个胸部协议的图像质量进行了定量和视觉评估。胸部和骨盆协议的视觉图像质量由放射科医生和放射技师使用视觉分级分析(VGA)方法进行评估。此外,使用CDRAD图像质量体模确定所有协议的定量图像质量数字逆()指标。此外,使用系统内置的剂量面积乘积(DAP)仪测量的DAP值评估胸部和骨盆协议的患者辐射剂量。在胸部和骨盆检查的各种协议设置中,实现了不同的定量图像质量和患者辐射剂量,但视觉图像质量评估显示,骨盆和胸部协议的所有观察者的图像质量均令人满意。与之前使用的特定系统协议相比,为所有成像系统协调胸部X射线摄影而选择的协议可降低患者的辐射暴露,同时保持足够的图像质量。胸部和骨盆X射线摄影的临床协议已根据患者辐射暴露和图像质量进行了标准化和优化。这种方法符合“尽可能合理地低”(ALARA)原则,在将辐射风险降至最低的同时确保获得最佳诊断信息。