Abstract

Introduction: Accurate and real-time monitoring of surgical blood loss is essential for ensuring intraoperative safety. However, there is currently no standard way to assess the amount of blood lost in patients during surgery. This study aims to evaluate the accuracy and precision of a new automatic intraoperative blood loss monitor, which can measure both free blood volume and blood content in sponges in real time. Methods: The monitor uses an integrated photoelectric probe to gauge hemoglobin levels in both free blood and blood taken from surgical sponges. These data, combined with initial hemoglobin levels, are processed using specific calculations to determine blood volume. We created 127 diverse free blood samples and 160 blood-containing sponge samples by using fresh pig blood and physiological saline. The monitor then measured these samples. We subsequently compared its measurements with actual values acquired through physical measurements, detecting both agreement and measurement errors. Repeated measurements were performed to calculate the coefficient of variation, thereby evaluating the monitor's precision. Results: The estimated blood loss percentage error of the monitor was 5.2% for free blood, -5.7% for small sponge, -6.3% for medium sponge, and -6.6% for large sponge. The coefficient of variation of free blood with different hemoglobin concentrations measured by the monitor was less than 10%. Bland-Altman analysis showed that the limits of agreement between the monitor and the reference method were all within the acceptable clinical range. Conclusion: The new automatic intraoperative blood loss monitor is an accurate and reliable device for monitoring both free blood and surgical sponge blood, and shows high performance under various clinical simulation conditions.

摘要

简介

准确、实时地监测手术失血量对于确保术中安全至关重要。然而,目前尚无评估手术期间患者失血量的标准方法。本研究旨在评估一种新型自动术中失血量监测仪的准确性和精密度,该监测仪可实时测量游离血容量和纱布中的血容量。

方法

该监测仪使用集成光电探头测量游离血和手术纱布中的血红蛋白水平。这些数据与初始血红蛋白水平结合,通过特定的计算处理,以确定血容量。我们使用新鲜猪血和生理盐水创建了 127 个不同的游离血样本和 160 个含血纱布样本。然后,监测仪对这些样本进行了测量。随后,我们将其测量值与通过物理测量获得的实际值进行了比较,以检测一致性和测量误差。我们还进行了重复测量,以计算变异系数,从而评估监测仪的精密度。

结果

监测仪对游离血的估计失血量百分比误差为 5.2%,对小纱布的误差为-5.7%,对中纱布的误差为-6.3%,对大纱布的误差为-6.6%。监测仪测量不同血红蛋白浓度的游离血的变异系数均小于 10%。Bland-Altman 分析显示,监测仪与参考方法之间的一致性界限均在可接受的临床范围内。

结论

新型自动术中失血量监测仪是一种准确可靠的监测游离血和手术纱布血的设备,在各种临床模拟条件下均表现出良好的性能。