Study of arterial blood pressure in newborn foals using an electronic sphygmomanometer.

作者信息

Franco R M, Ousey J C, Cash R S, Rossdale P D, Silver M

出版信息

Equine Vet J. 1986 Nov;18(6):475-8. doi: 10.1111/j.2042-3306.1986.tb03695.x.

Abstract

An electronic sphygmomanometer which functions on oscillometric principles, was used to measure arterial blood pressure in foals. Its accuracy was assessed by comparison with results obtained by direct measurement in anaesthetised pony foals, aged 34 to 64 days. Correlations between data obtained by the two methods were highly significant but the sphygmomanometric readings were consistently lower than those obtained by direct measurement. An equation was derived from the pooled values for mean, systolic and diastolic pressure measurements. This was used as a correction factor when the oscillometric method was employed in the field on Thoroughbred foals, during the first three days post partum, and on pony foals during the first day only. There were no significant differences between systolic, diastolic and mean values in Thoroughbreds on Days 1, 2 or 3 after birth. Measurements recorded on the first day after foaling were significantly lower (P less than 0.025) in ponies than in Thoroughbreds. Blood pressure measurements in Thoroughbreds and ponies were substantially higher than those reported previously using the ultrasonic Doppler technique.