Giguère Steeve, Knowles Harvey A, Valverde Alexander, Bucki Eric, Young Linda
Department of Large Animal Clinical Sciences, College of Veterinary Medicine University of Florida, Gainesville, FL 32610, USA.
J Vet Intern Med. 2005 Jul-Aug;19(4):571-6.
The objectives of this study were to assess, in anesthetized neonatal foals, the accuracy of 2 automated indirect oscillometric monitors for measurement of mean arterial pressure (MAP), to determine the optimal site of cuff placement for MAP monitoring, and to determine the relationship between arterial blood pressure and cardiac output. Ten neonatal foals were anesthetized and instrumented with a catheter in the metatarsal artery for direct MAP monitoring and measurement of cardiac output by lithium dilution. Concurrent MAP measurements were obtained with Cardell and Dinamap oscillometric monitors with cuffs placed at 3 different sites (coccygeal, metatarsal, and median arteries). Blood pressure was manipulated by varying the depth of anesthesia and by administration of dobutamine or phenylephrine. A statistically significant (P = .025) interaction was found between the type of monitor and cuff placement site. With the Cardell monitor, placement of the cuff over the coccygeal artery resulted in a significantly lower bias than placement over the median or dorsal metatarsal artery (P < .0001 and P = .0149, respectively). No significant difference in bias was found with cuff placement site when using the Dinamap monitor. The correlation coefficient (r) between MAP and cardiac output was 0.47. Indirect oscillometry with a cuff placed over the coccygeal artery or dorsal metatarsal artery is an acceptable method for measuring MAP in foals. Blood pressure does not correlate well with cardiac output in anesthetized foals.
本研究的目的是在麻醉的新生马驹中,评估2种自动间接示波监测仪测量平均动脉压(MAP)的准确性,确定用于MAP监测的袖带最佳放置部位,并确定动脉血压与心输出量之间的关系。10只新生马驹被麻醉,并在跖动脉插入导管以直接监测MAP并通过锂稀释法测量心输出量。使用Cardell和Dinamap示波监测仪在3个不同部位(尾动脉、跖动脉和正中动脉)放置袖带同时测量MAP。通过改变麻醉深度以及给予多巴酚丁胺或去氧肾上腺素来调节血压。发现监测仪类型与袖带放置部位之间存在统计学显著(P = 0.025)的交互作用。使用Cardell监测仪时,将袖带置于尾动脉上导致的偏差显著低于置于正中动脉或跖背动脉上(分别为P < 0.0001和P = 0.0149)。使用Dinamap监测仪时,袖带放置部位的偏差未发现显著差异。MAP与心输出量之间的相关系数(r)为0.47。在尾动脉或跖背动脉上放置袖带进行间接示波法是测量马驹MAP的可接受方法。在麻醉的马驹中,血压与心输出量的相关性不佳。