柠檬酸盐真空采血管采集前的质量评估——紫外分子吸收光谱法与离子色谱法的比较。

Preanalytical Quality Evaluation of Citrate Evacuated Blood Collection Tubes-Ultraviolet Molecular Absorption Spectrometry Confronted with Ion Chromatography.

机构信息

Faculty of Chemistry and Chemical Technology, University of Ljubljana, Večna Pot 113, SI1000 Ljubljana, Slovenia.

出版信息

Molecules. 2023 Nov 23;28(23):7735. doi: 10.3390/molecules28237735.

Abstract

We previously enabled a direct insight into the quality of citrate anticoagulant tubes before their intended use for specimen collection by introducing an easy-to-perform UV spectrometric method for citrate determination on a purified water model. The results revealed differences between the tubes of three producers, Greiner BIO-ONE (A), LT Burnik (B), and BD (C). It became apparent that tubes C contain an additive, which absorbs light in the ultraviolet range and prevents reliable evaluation of citrate anticoagulant concentration with the suggested method. In this research, we re-evaluate the quality of citrate-evacuated blood collection tubes by complementing UV spectrometry with ion chromatography. (1) Comparable results were obtained for tubes B at 220 nm. (2) Citrate concentrations determined with ion chromatography were lower for tubes A and C. Chromatograms reveal additional peaks for both. (3) Influences of heparin on absorption spectra and chromatograms of citrate were studied. Some similarities with the shape of the anticoagulant spectra of tubes A and C were observed, and the lithium heparin peak in chromatograms is close to them, but a confident judgment was not possible. (4) Contamination of anticoagulant solution with potassium, magnesium, and calcium was confirmed for all the brands, and contamination with lithium for B and C.

摘要

我们之前通过引入一种易于执行的 UV 光谱法,在纯化水模型上对柠檬酸盐进行测定,从而可以在使用前直接了解柠檬酸盐抗凝管的质量。结果表明,来自三个生产商的管之间存在差异:Greiner BIO-ONE(A)、LT Burnik(B)和 BD(C)。很明显,管 C 含有一种添加剂,它在紫外光范围内吸收光线,从而使我们无法使用建议的方法来可靠地评估柠檬酸盐抗凝剂浓度。在这项研究中,我们通过离子色谱法补充 UV 光谱法,重新评估了柠檬酸盐真空采血管的质量。(1)在 220nm 处,管 B 的结果相当。(2)离子色谱法测定的柠檬酸盐浓度对于管 A 和 C 较低。色谱图显示两者均有额外的峰。(3)研究了肝素对柠檬酸盐吸收光谱和色谱图的影响。观察到与管 A 和 C 的抗凝剂光谱形状有一些相似之处,并且色谱图中的肝素锂峰与之接近,但无法做出明确的判断。(4)证实所有品牌的抗凝剂溶液中都存在钾、镁和钙的污染,而 B 和 C 中存在锂的污染。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/16c3/10707995/a39ab5c16d35/molecules-28-07735-g001.jpg