Reekie R M, Carter K B, Owen H
Anaesthesia. 1986 Dec;41(12):1236-9. doi: 10.1111/j.1365-2044.1986.tb13010.x.
The performance of the Travenol infusor when used to administer an intravenous infusion of analgesic after surgery has been investigated. Mean flow was very similar to the manufacturer's quoted figure of 2 ml/hour, but there was a large interindividual variation. Results from a laboratory study suggest that this was due to variation in the manufacture of the infusor.
已经对特拉文诺输液泵在术后用于静脉输注镇痛药时的性能进行了研究。平均流速与制造商引用的2毫升/小时的数值非常相似,但个体间存在很大差异。一项实验室研究的结果表明,这是由于输液泵制造过程中的差异所致。