全髋关节置换术后5年假体松动——McKee-Farrar和Charnley假体的影像学研究

Loosening 5 years after total hip replacement--a radiological study of the McKee-Farrar and Charnley prostheses.

作者信息

Djerf K, Wahlström O, Hammerby S

出版信息

Arch Orthop Trauma Surg (1978). 1986;105(6):339-42. doi: 10.1007/BF00449939.

Abstract

In a prospective study, 138 patients who underwent total hip replacement by the McKee-Farrar or the Charnley technique were monitored yearly for 5 years by X-ray. Six variables were measured on anteroposterior and lateral projections to describe the position of the prosthesis. Lateral opening greater than 50 degrees and cover of the acetabular cup less than 1.5 cm correlated with an increased cup-loosening rate: 71% (98 of 138) were loose at the 5-year follow-up. However, only about one-third of all cup loosenings in the present series can be explained by nonoptimal position of the prosthesis. No significant difference emerged concerning loosening rate between the McKee-Farrar and the Charnley prostheses.

摘要

在一项前瞻性研究中,对138例行麦基-法拉尔(McKee-Farrar)或查理(Charnley)技术全髋关节置换术的患者进行了为期5年的每年一次的X线监测。在前后位和侧位片上测量六个变量以描述假体位置。外侧开口大于50度且髋臼杯覆盖小于1.5厘米与髋臼杯松动率增加相关:在5年随访时,71%(138例中的98例)出现松动。然而,在本系列中,所有髋臼杯松动中只有约三分之一可归因于假体位置欠佳。麦基-法拉尔假体和查理假体在松动率方面没有显著差异。