Stabilization of Traumatic Iliosacral Instability Using Innovative Implants: A Biomechanical Comparison.

作者信息

Grüneweller Niklas, Leunig Julia, Zderic Ivan, Gueorguiev Boyko, Wähnert Dirk, Vordemvenne Thomas

机构信息

Bielefeld University, Medical School and University Medical Center OWL, Protestant Hospital of the Bethel Foundation, Department of Trauma and Orthopedic Surgery, Burgsteig 13, 33617 Bielefeld, Germany.

AO Research Institute Davos, Clavadelerstrasse 8, 7270 Davos, Switzerland.

出版信息

J Clin Med. 2023 Dec 29;13(1):194. doi: 10.3390/jcm13010194.

Abstract

(1) Background: Demographic changes over the past decade have had a significant impact on pelvic ring fractures. They have increased dramatically in the orthogeriatric population. Surgeons are faced with implant fixation issues in the treatment of these fragility fractures. This study compares two innovative implants for stabilizing the iliosacral joint in a biomechanical setting. (2) Methods: An iliosacral screw with a preassembled plate allowing the placement of an additional short, angular stable screw in the ilium and a triangular fixation system consisting of a fenestrated ilium screw and an iliosacral screw quasi-statically inserted through the "fenestra" were instrumented in osteoporotic artificial bone models with a simulated Denis zone 1 fracture. Biomechanical testing was performed on a servo-hydraulic testing machine using increasing, synchronous axial and torsional sinusoidal cyclic loading to failure. (3) Results: The SI-Plate and TriFix showed comparable stiffness values. The values for fracture gap angle and screw tip cutout were significantly lower for the TriFix compared to the SI-Plate. In addition, the number of cycles to failure was significantly higher for the TriFix. (4) Conclusions: Implant anchorage and primary stability can be improved in iliosacral instability using the triangular stabilization system.

摘要

(1) 背景:过去十年的人口结构变化对骨盆环骨折产生了重大影响。在老年骨科人群中,此类骨折显著增加。外科医生在治疗这些脆性骨折时面临植入物固定问题。本研究在生物力学环境下比较了两种用于稳定骶髂关节的创新型植入物。(2) 方法:在模拟Denis 1区骨折的骨质疏松人工骨模型中植入一种带有预组装钢板的骶髂螺钉,该钢板允许在髂骨中额外放置一枚短的角稳定螺钉,以及一种由带孔髂骨螺钉和通过“窗口”准静态插入的骶髂螺钉组成的三角形固定系统。在伺服液压试验机上进行生物力学测试,使用递增的同步轴向和扭转正弦循环加载直至失效。(3) 结果:SI-Plate和TriFix显示出可比的刚度值。与SI-Plate相比,TriFix的骨折间隙角度和螺钉尖端穿出值显著更低。此外,TriFix的失效循环次数显著更高。(4) 结论:使用三角形稳定系统可改善骶髂关节不稳定时的植入物锚固和初始稳定性。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/422d/10779522/f9fb6b05d42a/jcm-13-00194-g001.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验