Dosimetric Comparison of Unmatched Flattening Filter-free and Flattened Beams in Volumetric Arc Therapy Plans for Head-and-neck Cancer.

作者信息

Kumar Arvind, Sharma Kiran, Bhatt Chandi Prasad, Garg Abhishek

机构信息

Department of Physics, Graphic Era (Deemed to be University), Dehradun, India.

Department of Radiation Oncology, All India Institute of Medical Sciences, Rishikesh, Uttarakhand, India.

出版信息

J Med Phys. 2023 Oct-Dec;48(4):338-344. doi: 10.4103/jmp.jmp_68_23. Epub 2023 Dec 5.

Abstract

INTRODUCTION

Conventionally, the fattened beam is being used in radiotherapy for routine clinical cases even after introduction of intensity-modulated radiotherapy with incorporation of multi-leaf collimator system. With the removal of the flattening filter (FF) from the beam's path average energy of the photon gets reduced resulting in reduced scatter, reduction in treatment time, and reduced neutron contamination for high-energy beam and ultimately resulting in treatment plan quality deviations. This study aims to investigate the usefulness of the FF-free (FFF) beam for routine head-and-neck cancer (HNC) cases treated with volumetric arc therapy (VMAT) and dosimetrically compares the result with the FF beam.

MATERIALS AND METHODS

In this study, 20 patients treated on Halcyon (unmatched 6 megavoltage [6MV] FFF beam) medical linear accelerator with VMAT of different HNC selected and for comparison with 6MV FF beam, 20 equivalent treatment plans are created for Truebeam configuration and the plans have been evaluated for target coverage, doses to the organ at risk (OAR), and other dose quality indices.

RESULTS

Comparable target coverage, doses to OARs except for rest right parotid ( = 0.02) between 6MV FFF beam and 6MV FF beam is observed. Insignificant differences in conformity index, homogeneity index, and gradient index have been observed. Higher monitor unit (MU) ( ≤ 0.001) and lesser beam on time (BOT) ( = 0.003) have been observed in 6MV FFF.

CONCLUSION

6MV FFF beam provides comparable target coverage and improved dose-sparing effect to most of the OARs. 6MV FFF beam has lesser BOT, but on the other hand number of MUs is higher as compared to 6MV-FF plans.

摘要

引言

传统上,即使在引入了包含多叶准直器系统的调强放射治疗之后,在常规临床病例的放射治疗中仍使用均整后的射束。随着射束路径中的均整滤过器(FF)被移除,光子的平均能量降低,从而导致散射减少、治疗时间缩短以及高能射束的中子污染减少,最终导致治疗计划质量出现偏差。本研究旨在探讨无均整滤过器(FFF)射束在容积弧形调强放疗(VMAT)治疗的常规头颈癌(HNC)病例中的实用性,并在剂量学方面将结果与均整滤过器射束进行比较。

材料与方法

在本研究中,选取了20例在Halcyon(非匹配的6兆伏[6MV]FFF射束)医用直线加速器上接受VMAT治疗的不同HNC患者,为了与6MV均整滤过器射束进行比较,针对Truebeam配置创建了20个等效治疗计划,并对这些计划的靶区覆盖情况、危及器官(OAR)的剂量以及其他剂量质量指标进行了评估。

结果

观察到6MV FFF射束和6MV均整滤过器射束之间在靶区覆盖情况、除右侧腮腺外的OAR剂量方面具有可比性(P = 0.02)。在适形指数、均匀性指数和梯度指数方面观察到无显著差异。在6MV FFF射束中观察到更高的监测单位(MU)(P ≤ 0.001)和更短的射束开启时间(BOT)(P = 0.003)。

结论

6MV FFF射束提供了可比的靶区覆盖,并对大多数OAR具有更好的剂量 sparing效应。6MV FFF射束的BOT更短,但另一方面,与6MV均整滤过器计划相比,MU数量更高。