Suppr超能文献

反式全肩关节置换术中肱骨短柄的内翻-外翻对线:真的有关系吗?

Varus-valgus alignment of humeral short stem in reverse total shoulder arthroplasty: does it really matter?

机构信息

Department of Orthopedic Surgery, Seoul St. Mary's Hospital, The Catholic University of Korea, Seoul, Republic of Korea.

Orthopaedic Biomechanics Laboratory, Congress Medical Foundation, Pasadena, CA, USA.

出版信息

J Shoulder Elbow Surg. 2024 Aug;33(8):1762-1770. doi: 10.1016/j.jse.2023.11.024. Epub 2024 Jan 17.

Abstract

BACKGROUND

The utilization of short humeral stems in reverse total shoulder arthroplasty has gained attention in recent times. However, concerns regarding the risk of misalignment during implant insertion are associated with their use.

METHODS

Eight fresh-frozen cadaveric shoulders were prepared for dissection and biomechanical testing. A bespoke humeral implant was fabricated to facilitate assessment of neutral, varus, and valgus alignments using a single stem, and 10° was established as the maximum permissible angle for misalignments. Shift in humerus position and changes in deltoid length attributable to misalignments relative to the neutral position were evaluated using a Microscribe 3DLx system. The impingement-free range of motion, encompassing abduction, adduction, internal rotation, and external rotation (ER), was gauged using a digital goniometer. The capacity for abduction was evaluated at maximal abduction angles under successive loading on the middle deltoid. A specialized traction system coupled with a force transducer was employed to measure anterior dislocation forces.

RESULTS

Relative to the neutral alignment, valgus alignment resulted in a more distal (10.5 ± 2.4 mm) and medial (8.3 ± 2.2 mm) translation of the humeral component, whereas the varus alignment resulted in the humerus shifting more superiorly (11.2 ± 1.3 mm) and laterally (9.9 ± 0.9 mm) at 0° abduction. The valgus alignment exhibited the highest abduction angle than neutral alignment (86.2°, P < .001). Conversely, the varus alignment demonstrated significantly higher adduction (18.4 ± 7.4°, P < .001), internal rotation (68.9 ± 15.0°, P = .014), and ER (45.2 ± 10.5°, P = .002) at 0° abduction compared to the neutral alignments. Anterior dislocation forces were considerably lower (23.8 N) in the varus group compared to the neutral group at 0°ER (P = .047). Additionally, abduction capability was markedly higher in varus alignment at low deltoid loads than the neutral alignment (5N, P = .009; 7.5 N, P = .007).

CONCLUSIONS

The varus position enhances rotational range of motion (ROM) but increases instability, while the valgus position does not significantly impact ROM or instability compared to the neutral position.

摘要

背景

在反式全肩关节置换术中,使用短肱骨柄已经引起了关注。然而,由于在植入物插入过程中存在错位的风险,因此人们对其使用存在担忧。

方法

准备了 8 个新鲜冷冻的尸体肩部进行解剖和生物力学测试。定制了肱骨植入物,以使用单个柄评估中立、内翻和外翻对齐的情况,并且将 10°确定为允许的最大错位角度。使用 Microscribe 3DLx 系统评估相对于中立位置的肱骨位置和三角肌长度的变化。使用数字测角计测量无撞击的运动范围,包括外展、内收、内旋和外旋(ER)。在中三角肌的连续加载下,以最大外展角度评估外展能力。使用特殊的牵引系统和力传感器测量前脱位力。

结果

与中立对齐相比,外翻对齐导致肱骨组件更向远端(10.5±2.4mm)和内侧(8.3±2.2mm)平移,而内翻对齐导致肱骨更向上方(11.2±1.3mm)和外侧(9.9±0.9mm)平移,在 0°外展时。外翻对齐的外展角度比中立对齐高(86.2°,P<.001)。相反,与中立对齐相比,内翻对齐在 0°外展时显示出更高的内收(18.4±7.4°,P<.001)、内旋(68.9±15.0°,P=.014)和 ER(45.2±10.5°,P=.002)。在 0°ER 时,与中立组相比,内翻组的前脱位力明显较低(23.8N,P=.047)。此外,与中立对齐相比,在低三角肌负荷下,内翻对齐的外展能力明显更高(5N,P=.009;7.5N,P=.007)。

结论

与中立位置相比,内翻位置增加了旋转活动范围(ROM)但增加了不稳定性,而外翻位置对 ROM 或不稳定性没有明显影响。