Ejection fraction and ventricular volumes on rubidium positron emission tomography: Validation against cardiovascular magnetic resonance.

机构信息

Houston Methodist DeBakey Heart & Vascular Center, Houston, TX, USA.

Division of Cardiovascular Medicine, University of Michigan, Ann Arbor, MI, USA.

出版信息

J Nucl Cardiol. 2024 Feb;32:101810. doi: 10.1016/j.nuclcard.2024.101810. Epub 2024 Jan 28.

Abstract

BACKGROUND

Cardiovascular magnetic resonance (CMR) is the non-invasive gold standard for non-invasively determining left ventricular volumes (LVVs) and ejection fraction (EF). We aimed to assess the accuracy of LVV and left ventricular ejection fraction measured by positron emission tomography (PET) as compared to CMR.

METHODS

Patients who underwent both PET and CMR within 1 year were identified from prospective institutional registries. Analysis was performed to evaluate the agreement between the raw and body-surface-area-normalized left ventricular volume (LVV) and EF derived from PET vs. those derived from CMR.

RESULTS

The study population consisted of 669 patients (mean age 62 ± 13 years, 65% male). The median (interquartile range [IQR]) duration between CMR and PET imaging was 36 (7-118) days. The median (IQR) EF values were 52% (38-63%) on CMR and 53% (37-65%) on PET (mean difference: 0.53% ± 9.1, P = 0.129) with a strong correlation (Spearman rho = 0.84, P < 0.001; Intraclass Correlation Coefficient 0.84, 95% confidence interval [CI]: 0.82-0.86, P < 0.001; Lin's concordance correlation coefficient was 0.844, 95% CI: 0.822 to 0.865). Results were similar with LVV, normalized LVV/EF, and in subgroups of patients with reduced EF, coronary artery disease scar, and LV hypertrophy as well as in patients with defibrillators. However, PET tended to underestimate LVV compared to CMR.

CONCLUSION

Our analysis showed a strong correlation of EF and LVV by PET against a reference standard of CMR, whereas PET significantly underestimated LVV, but not EF, compared to CMR.

摘要

背景

心血管磁共振(CMR)是无创性确定左心室容积(LVV)和射血分数(EF)的金标准。我们旨在评估正电子发射断层扫描(PET)测量的 LVV 和左心室射血分数与 CMR 相比的准确性。

方法

从前瞻性机构登记处确定了在 1 年内同时接受 PET 和 CMR 的患者。进行分析以评估从 PET 获得的未经处理和基于体表面积归一化的左心室容积(LVV)和 EF 与从 CMR 获得的那些之间的一致性。

结果

研究人群包括 669 名患者(平均年龄 62 ± 13 岁,65%为男性)。CMR 和 PET 成像之间的中位(四分位距 [IQR])时间为 36(7-118)天。CMR 上的中位(IQR)EF 值为 52%(38-63%),PET 上为 53%(37-65%)(平均差异:0.53% ± 9.1,P = 0.129),具有很强的相关性(Spearman rho = 0.84,P < 0.001;组内相关系数 0.84,95%置信区间 [CI]:0.82-0.86,P < 0.001;Lin 的一致性相关系数为 0.844,95%CI:0.822 至 0.865)。结果与 LVV、归一化 LVV/EF 以及 EF 降低、冠状动脉疾病疤痕、LV 肥厚以及装有除颤器的患者亚组相似。然而,与 CMR 相比,PET 往往会低估 LVV。

结论

我们的分析显示,EF 和 LVV 通过 PET 与 CMR 的参考标准具有很强的相关性,而 PET 与 CMR 相比显著低估了 LVV,但 EF 则没有。