植入式环路记录仪监测随访的患者中假阳性警报:系统评价。

False-positive alarms in patients with implantable loop recorder followed by remote monitoring: A systematic review.

机构信息

Cardiology Unit, Department of Medical Translational Science, University of Campania "Luigi Vanvitelli"-Monaldi Hospital, Naples, Italy.

出版信息

Pacing Clin Electrophysiol. 2024 Mar;47(3):406-416. doi: 10.1111/pace.14941. Epub 2024 Feb 11.

Abstract

Remote Monitoring (RM) has been shown to provide useful information about arrhythmic events in patients with implantable loop recorders (ILRs), however there is few and conflicting data about the false positive (FP) alarms burden and characteristics among ILR recipients. The aim of the present systematic review was to evaluate incidence and characteristics of FP alarms among ILR patients followed by RM. We developed a systematic research in Embase, MEDLINE and PubMed databases and selected all papers focused on false positive ILR transmissions published from June 1, 2013 to June 1, 2023. Case reports, meeting summaries, posters and simple reviews were excluded. Twelve reports were finally selected, including five prospective and seven retrospective studies. Information about population characteristics, device type and setting, overall transmissions and FP alarms and any adopted strategies to reduce them were extracted from an overall population of 3.305 patients. FP alarms were 59.7% of the overall remote transmissions and were found in 1/5 of the analyzed population. FP alarms for atrial fibrillation were the most common cause of false transmissions and were mainly due to premature atrial and ventricular complexes. No clinical predictors of FP alarms were identified, except for nonparasternal ILR implantation site. Since the overload work due to FP alarms might reduce the benefit of remote monitoring of ILR patients, the device optimization is an important step until an help from machine-learning algorithms is available.

摘要

远程监测(RM)已被证明可以提供有关植入式环路记录器(ILR)患者心律失常事件的有用信息,但是关于 ILR 受者中假阳性(FP)警报负担和特征的数据很少且相互矛盾。本系统评价的目的是评估 RM 后 ILR 患者中 FP 警报的发生率和特征。我们在 Embase、MEDLINE 和 PubMed 数据库中进行了系统研究,并选择了所有关注 2013 年 6 月 1 日至 2023 年 6 月 1 日期间发布的关于 ILR 传输假阳性的论文。排除病例报告、会议摘要、海报和简单评论。最终选择了 12 份报告,其中包括 5 项前瞻性研究和 7 项回顾性研究。从 3305 名患者的总体人群中提取了有关人口统计学特征、设备类型和设置、总体传输和 FP 警报以及任何用于减少 FP 警报的策略的信息。FP 警报占总远程传输的 59.7%,在分析人群的 1/5 中发现。房颤的 FP 警报是假传输最常见的原因,主要是由于房性和室性早搏。除了非心前区 ILR 植入部位外,未发现 FP 警报的临床预测因素。由于 FP 警报导致的工作负荷过重可能会降低 ILR 患者远程监测的益处,因此设备优化是一个重要步骤,直到可以使用机器学习算法提供帮助。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验