Abstract

The National Committee for Clinical Laboratory Standards (NCCLS) has published a tentative standard for leukocyte differential counting, by means of which a manual or automated method for leukocyte differential counting can be compared with a manual reference method. The performance of the Technicon H6000 system was evaluated using the standard at Stamford and Overlook Hospitals. A total of 502 patient samples were analyzed: 315 from Overlook and 187 from Stamford. The H6000 system was found to be approximately four times more precise than the 200-cell manual reference method for each cell type. Correlation of the H6000 system with the manual method was good, with correlation coefficients of 0.98 for neutrophils and lymphocytes, 0.96 for eosinophils, 0.72 for monocytes, and 0.5 for basophils. The clinical sensitivity of the H6000 system, measured in terms of false normals and false abnormals, was similar to that of the manual reference method when measured against itself. There were no clinically significant discrepancies in results from the H6000 system, except for possibly one case where a patient was already on antibiotic therapy. The NCCLS standard was found to be a useful but rather complex and involved method for evaluating the performance of the H6000 system, the major problem being the amount of work needed to count manually the number of cells required for the manual reference method.

摘要

美国国家临床实验室标准委员会(NCCLS)已发布了一份白细胞分类计数的暂行标准,借此可将白细胞分类计数的手工或自动化方法与手工参考方法进行比较。在斯坦福医院和俯瞰医院使用该标准对Technicon H6000系统的性能进行了评估。共分析了502份患者样本:315份来自俯瞰医院,187份来自斯坦福医院。结果发现,对于每种细胞类型,H6000系统的精确性约为200细胞手工参考方法的四倍。H6000系统与手工方法的相关性良好,中性粒细胞和淋巴细胞的相关系数为0.98,嗜酸性粒细胞为0.96,单核细胞为0.72,嗜碱性粒细胞为0.5。以假正常和假异常衡量,H6000系统的临床敏感性与手工参考方法自身对照时相似。H6000系统的结果中不存在临床上显著的差异,可能除了1例患者正在接受抗生素治疗的情况。发现NCCLS标准是评估H6000系统性能的一种有用但相当复杂且繁琐的方法,主要问题是手工计数手工参考方法所需细胞数量的工作量。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验