Evaluation of impedance technique for fluid-volume monitoring during hemodialysis.

作者信息

Tedner B, Lins L E, Asaba H, Wehle B

出版信息

Int J Clin Monit Comput. 1985;2(1):3-8. doi: 10.1007/BF02915866.

Abstract

Fluid-volume changes during hemodialysis were studied in 30 patients at four separate treatments in each patient. Readings of fluid-volume changes obtained by a fluid-balance monitor with non-invasive tetrapolar impedance technique were compared to changes in weight caused by the treatment. A correlation coefficient of r = 0.90 (p less than 0.001) was found. A survey of papers describing impedance measurements during hemodialysis or diuresis is also presented.

摘要