Suppr超能文献

采用阻抗技术监测血液透析过程中的液体量变化。

Fluid volume changes during hemodialysis monitored with the impedance technique.

作者信息

Tedner B, Lins L E

出版信息

Artif Organs. 1985 Nov;9(4):416-20. doi: 10.1111/j.1525-1594.1985.tb04403.x.

Abstract

A fluid monitor based on the tetrapolar impedance technique was employed for continuous measurements of fluid volume change during hemodialysis treatment and compared to concomitant ultrafiltration and bed scale readings. Oral fluid intake before and during the dialysis immediately could be recorded with the bed scale as a weight change. In the readings of the fluid monitor, however, the response to fluid intake could be recorded only gradually, as the fluid was absorbed from the gastrointestinal tract. Taking this into account, together with a previously described artifact due to shifts in body position, an excellent agreement was obtained between impedance and bed scale fluid volume monitoring. It is concluded that impedance measurements give a better estimation of the amount of metabolically active fluid than does body weight monitoring.

摘要

采用基于四极阻抗技术的液体监测仪,在血液透析治疗期间连续测量液体体积变化,并与同时进行的超滤和床旁秤读数进行比较。透析前和透析过程中的口服液体摄入量可通过床旁秤记录体重变化来确定。然而,在液体监测仪的读数中,由于液体从胃肠道吸收,对液体摄入的反应只能逐渐记录。考虑到这一点,再加上先前描述的因体位改变引起的伪影,阻抗和床旁秤液体体积监测之间取得了极好的一致性。得出的结论是,与体重监测相比,阻抗测量能更好地估计代谢活性液体的量。