Brind J L, Kuo S W, Chervinsky K, Fitzgerald K, Orentreich N
Steroids. 1985 Jun;45(6):565-71. doi: 10.1016/0039-128x(85)90021-2.
TLC plates with a 25 mu thick polyamide stationary phase were modified for the separation of neutral steroids by impregnation with propylene glycol. A mixture of tritiated 5 alpha-androstan-3 alpha,17 beta-diol, testosterone, 17 beta-hydroxy-5 alpha-androstan-3-one and 4-androstene-3,17-dione was applied to the plate and developed in a toluene mobile phase to a height of 13.6 cm. This resulted in complete resolution of the 4 compounds as detected by a gas flow scanner or imaging analyzer. Cutting and elution of peak areas with methanol resulted in quantitative recovery of all four steroids. The thinness of the layer also permitted a 3-5% counting efficiency on scanning, resulting in good quantitation of recovery without liquid scintillation counting. The high sorptive capacity of the polyamide layer also enabled extracts of normal human serum to be defatted on the TLC plate by development with pure hexane prior to the toluene step. The new method thus offers several advantages over existing methods for steroid separations and should be adaptable to separations of other relatively non-polar compounds.
使用厚度为25微米的聚酰胺固定相薄层层析板,通过用丙二醇浸渍对其进行改性,以用于中性甾体的分离。将含有氚标记的5α-雄甾烷-3α,17β-二醇、睾酮、17β-羟基-5α-雄甾烷-3-酮和4-雄烯-3,17-二酮的混合物点样到薄板上,并在甲苯流动相中展开至13.6厘米的高度。通过气流扫描仪或成像分析仪检测,这使得这4种化合物得到了完全分离。用甲醇切割并洗脱峰面积,实现了所有4种甾体的定量回收。薄层的薄度还使得扫描时计数效率达到3%-5%,从而在无需液体闪烁计数的情况下实现了良好的回收定量。聚酰胺层的高吸附能力还使得正常人血清提取物在甲苯展开步骤之前,通过用纯己烷展开在薄层层析板上脱脂。因此,这种新方法相对于现有的甾体分离方法具有多个优点,并且应该适用于其他相对非极性化合物的分离。