Suppr超能文献

Sysmex XN血液分析仪检测髓系肿瘤中未成熟粒细胞参数的评估

Evaluation of immature granulocyte parameters in myeloid neoplasms assayed by Sysmex XN hematology analyzer.

作者信息

Lu Qifeng, Li Ying, Li Tian, Hou Tingting, Zhao Yajuan, Feng Shu, Yang Xixian, Zhu Mengyu, Shen Yajuan

机构信息

Department of Clinical Laboratory, Shandong Provincial Hospital Affiliated to Shandong First Medical University, Jinan, 250021, Shandong Province, China.

Department of Hematology, Shandong Provincial Hospital Affiliated to Shandong First Medical University, Jinan, 250021, Shandong Province, China.

出版信息

J Hematop. 2022 Mar;15(1):1-6. doi: 10.1007/s12308-022-00484-w. Epub 2022 Feb 8.

Abstract

Immature granulocytes (IGs) have significance for the diagnosis of myeloid neoplasms (MNs). The current study aims to use a hematology analyzer to evaluate the accuracy of IG parameters in MNs. Blood specimens from 388 patients with MN, 524 with non-hematological neoplasms (non-HNs), including 109 patients with inflammation and 68 undergoing G-CSF administration, and 500 healthy control subjects were analyzed. IG parameters was assayed by Sysmex XN-9000 (XN) and compared with manual assessments. A high level of agreement between IG% derived from XN and manual measurements for MN patients (r = 0.828, p < 0.0001) was revealed but only a moderate correlation for acute myeloid leukemia patients (AML; r = 0.597; p < 0.0001). Bland-Altman bias analysis was conducted, and the results showed that differences in IG% from XN and manual analysis for MN patients were considered clinically insignificant. ROC analysis demonstrated a good performance of IG# (AUC = 0.842) and IG% (AUC = 0.885) assessed by XN for MN patients with cut-off values of 0.200 × 10/L and 1.95%, respectively. IG parameters from Sysmex XN analyzer are helpful for screening of MNs even though granulocyte morphological abnormalities may interfere with IG parameter accuracy.

摘要

未成熟粒细胞(IGs)对髓系肿瘤(MNs)的诊断具有重要意义。本研究旨在使用血液分析仪评估IG参数在MNs诊断中的准确性。分析了388例MN患者、524例非血液系统肿瘤(非HNs)患者(包括109例炎症患者和68例接受粒细胞集落刺激因子(G-CSF)治疗的患者)以及500例健康对照者的血液标本。采用Sysmex XN-9000(XN)检测IG参数,并与手工评估结果进行比较。结果显示,XN检测的IG%与MN患者手工测量结果高度一致(r = 0.828,p < 0.0001),但急性髓系白血病(AML)患者的相关性仅为中等(r = 0.597;p < 0.0001)。进行了布兰德-奥特曼偏差分析,结果表明XN检测与手工分析所得IG%的差异在临床上无显著意义。ROC分析显示,XN检测的IG#(曲线下面积(AUC)= 0.842)和IG%(AUC = 0.885)对MN患者具有良好的诊断性能,其截断值分别为0.200×10⁹/L和1.95%。尽管粒细胞形态异常可能会干扰IG参数的准确性,但Sysmex XN分析仪检测的IG参数有助于MNs的筛查。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/a5d3/10869398/5edf30b605c8/12308_2022_484_Fig1_HTML.jpg

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验