一种新型无线非粘性心肺监测设备对早产儿皮肤状况的影响。

The Effects of a New Wireless Non-Adhesive Cardiorespiratory Monitoring Device on the Skin Conditions of Preterm Infants.

机构信息

Máxima Medical Center, Department of Neonatology, De Run 4600, 5504 DB Veldhoven, The Netherlands.

Department of Mathematics and Computer Science, Eindhoven University of Technology, Groene Loper 3, 5612 AE Eindhoven, The Netherlands.

出版信息

Sensors (Basel). 2024 Feb 16;24(4):1258. doi: 10.3390/s24041258.

Abstract

AIM

The aim of our study was to investigate skin conditions when wearing and removing a novel wireless non-adhesive cardiorespiratory monitoring device for neonates (Bambi-Belt) compared to standard adhesive electrodes.

STUDY DESIGN

This was a prospective study including preterm neonates requiring cardiorespiratory monitoring. Besides standard electrodes, the infants wore a Bambi Belt for 10 consecutive days. Their skin conditions were assessed using Trans Epidermal Water Loss (TEWL) and the Neonatal Skin Condition Score (NSCS) after daily belt and standard electrode removal. The ∆TEWL was calculated as the difference between the TEWL at the device's location (Bambi-Belt/standard electrode) and the adjacent control skin location, with a higher ∆TEWL indicating skin damage.

RESULTS

A total of 15 infants (gestational age (GA): 24.1-35.6 wk) were analyzed. The ΔTEWL significantly increased directly after electrode removal (10.95 ± 9.98 g/m/h) compared to belt removal (5.18 ± 6.71 g/m/h; F: 8.73, = 0.004) and after the washout period (3.72 ± 5.46 g/m/h vs. 1.86 ± 3.35 g/m/h; F: 2.84, = 0.09), although the latter did not reach statistical significance. The TEWL was not influenced by prolonged belt wearing. No significant differences in the NSCS score were found between the belt and electrode (OR: 0.69, 95% CI [0.17, 2.88], = 0.6).

CONCLUSION

A new wireless non-adhesive device for neonatal cardiorespiratory monitoring was well tolerated in preterm infants and may be less damaging during prolonged wearing.

摘要

目的

本研究旨在比较新型无线非粘性新生儿心肺监测仪(Bambi-Belt)与标准粘性电极在佩戴和移除时对皮肤状况的影响。

研究设计

这是一项前瞻性研究,纳入需要心肺监测的早产儿。除了标准电极外,婴儿还连续佩戴 Bambi 带 10 天。每天移除腰带和标准电极后,使用经皮水分丢失(TEWL)和新生儿皮肤状况评分(NSCS)评估皮肤状况。TEWL 的差值(∆TEWL)为设备位置(Bambi 带/标准电极)与相邻对照皮肤位置的 TEWL 差值,∆TEWL 越高表明皮肤损伤越大。

结果

共分析了 15 名婴儿(胎龄(GA):24.1-35.6 周)。与腰带移除(5.18 ± 6.71 g/m/h)相比,电极移除后 TEWL 直接显著增加(10.95 ± 9.98 g/m/h;F:8.73, = 0.004),与冲洗期后(3.72 ± 5.46 g/m/h 比 1.86 ± 3.35 g/m/h;F:2.84, = 0.09)相比,虽然后者未达到统计学意义。长时间佩戴腰带不会影响 TEWL。Bambi 带和电极之间的 NSCS 评分无显著差异(OR:0.69,95%CI [0.17, 2.88], = 0.6)。

结论

新型无线非粘性新生儿心肺监测仪在早产儿中耐受性良好,长时间佩戴时可能对皮肤损伤较小。