在温暖条件下运动时,MX3 便携式汗液钠分析仪的可靠性和有效性。

Reliability and validity of the MX3 portable sweat sodium analyser during exercise in warm conditions.

机构信息

Research Institute for Sport and Exercise Science (UCRISE), University of Canberra, Bruce, Canberra, ACT, Australia.

出版信息

Eur J Appl Physiol. 2024 Jul;124(7):2153-2160. doi: 10.1007/s00421-024-05447-w. Epub 2024 Mar 2.

Abstract

PURPOSE

Accurately measuring sweat sodium concentration ([Na]) in the field is advantageous for coaches, scientists, and dieticians looking to tailor hydration strategies. The MX3 hydration testing system is a new portable analyser that uses pre-calibrated biosensors to measure sweat [Na]. This study aimed to assess the validity and reliability of the MX3 hydration testing system.

METHODS

Thirty-one (11 females) recreationally active participants completed one experimental trial. During this trial, participants exercised at a self-selected pace for 45 min in a warm environment (31.5 ± 0.8 °C, 63.2 ± 1.3% relative humidity). Sweat samples were collected from three measurement sites using absorbent patches. The samples were then analysed for sweat [Na] using both the MX3 hydration testing system and the Horiba LAQUAtwin-NA-11. The reliability of the MX3 hydration testing system was determined following two measurements of the same sweat sample.

RESULTS

The mean difference between measurements was 0.1 mmoL·L (95% limits of agreement (LoA): - 9.2, 9.4). The analyser demonstrated a coefficient of variation (CV) of 5.6% and the standard error of measurement was 3.3 mmoL·L. When compared to the Horiba LAQUAtwin-NA-11, there was a mean difference of - 1.7 mmoL·L (95% LoA: - 0.25 , 0.25 ) and the CV was 9.8%.

CONCLUSION

The MX3 hydration testing system demonstrated very good single-trial reliability, moderate agreement and a very good CV relative to the Horiba LAQUAtwin-Na-11. To further validate its performance, the MX3 hydration testing system should be compared with analytical techniques known for superior reliability and validity.

摘要

目的

在现场准确测量汗钠浓度([Na]),有利于教练、科学家和营养师制定个性化的水合策略。MX3 水合测试系统是一种新的便携式分析仪,它使用经过预校准的生物传感器来测量汗[Na]。本研究旨在评估 MX3 水合测试系统的有效性和可靠性。

方法

31 名(11 名女性)有运动习惯的参与者完成了一项实验性试验。在这项试验中,参与者在温暖的环境中(31.5±0.8°C,63.2±1.3%相对湿度)以自定速度运动 45 分钟。使用吸收垫从三个测量部位采集汗液样本。然后使用 MX3 水合测试系统和 Horiba LAQUAtwin-NA-11 对样本进行汗[Na]分析。通过对同一汗液样本进行两次测量,确定了 MX3 水合测试系统的可靠性。

结果

两次测量之间的平均差值为 0.1 mmoL·L(95% 置信区间(LoA):-9.2,9.4)。该分析仪的变异系数(CV)为 5.6%,测量标准差为 3.3 mmol·L。与 Horiba LAQUAtwin-NA-11 相比,差值平均为-1.7 mmoL·L(95% LoA:-0.25,0.25),CV 为 9.8%。

结论

与 Horiba LAQUAtwin-NA-11 相比,MX3 水合测试系统在单次试验中表现出非常好的可靠性、适度的一致性和非常好的 CV。为了进一步验证其性能,应将 MX3 水合测试系统与以可靠性和有效性著称的分析技术进行比较。