Peterson Kelly A, Francis Ryan M, Banach Catherine A, Bradley Ashley M, Burton Sarah D, Erickson Jeremy D, Lockwood Schuyler P, Jensen Karissa L, Yokosuk Michael O, Johnson Timothy J, Myers Tanya L
Appl Opt. 2024 Feb 20;63(6):1553-1565. doi: 10.1364/AO.514661.
Obtaining the complex refractive index vectors () and () allows calculation of the (infrared) reflectance spectrum that is obtained from a solid in any of its many morphological forms. We report an adaptation to the KBr pellet technique using two gravimetric dilutions to derive quantitative ()/() for dozens of powders with greater repeatability. The optical constants of bisphenol A and sucrose are compared to those derived by other methods, particularly for powdered materials. The variability of the values for bisphenol A was examined by 10 individual measurements, showing an average coefficient of variation for peak heights of 5.6%. Though no established standards exist, the pellet-derived peak values of bisphenol A differ by 11% and 31% from their single-angle- and ellipsometry-derived values, respectively. These values provide an initial estimate of the precision and accuracy of complex refractive indices that can be derived using this method. Limitations and advantages of the method are discussed, the salient advantage being a more rapid method to derive / for those species that do not readily form crystals or specular pellets.
获取复折射率矢量()和()能够计算从任何多种形态的固体中获得的(红外)反射光谱。我们报告了一种对溴化钾压片技术的改进方法,该方法使用两次重量稀释来推导数十种粉末的定量()/(),具有更高的重复性。将双酚A和蔗糖的光学常数与通过其他方法得出的光学常数进行比较,特别是对于粉末材料。通过10次单独测量检查了双酚A的 值的变异性,结果表明 峰高的平均变异系数为5.6%。尽管不存在既定标准,但通过压片法得出的双酚A的 峰值分别与其通过单角度法和椭偏法得出的值相差11%和31%。这些值提供了使用该方法可得出的复折射率的精度和准确性的初步估计。讨论了该方法的局限性和优点,其显著优点是对于那些不易形成晶体或镜面压片的物质,是一种更快速的推导 / 的方法。