Lee Soohyun, Choi Stacey S, Meleppat Ratheesh K, Zawadzki Robert J, Doble Nathan
College of Optometry, The Ohio State University, 338 West 10th Avenue, Columbus, Ohio 43210, USA.
Department of Ophthalmology and Visual Sciences, Havener Eye Institute, The Ohio State University, 915 Olentangy River Road, Suite 5000, Ohio 43212, USA.
Biomed Opt Express. 2024 Feb 22;15(3):1815-1830. doi: 10.1364/BOE.507449. eCollection 2024 Mar 1.
High-speed, phase contrast retinal and blood flow imaging using an adaptive optics partially confocal multi-line ophthalmosocope (AO-pcMLO) is described. It allows for simultaneous confocal and phase contrast imaging with various directional multi-line illumination by using a single 2D camera and a digital micromirror device (DMD). Both vertical and horizontal line illumination directions were tested, for photoreceptor and vascular imaging. The phase contrast imaging provided improved visualization of retinal structures such as cone inner segments, vessel walls and red blood cells with images being acquired at frame rates up to 500 Hz. Blood flow velocities of small vessels (<40 m in diameter) were measured using kymographs for capillaries and cross-correlation between subsequent images for arterioles or venules. Cardiac-related pulsatile patterns were observed with normal resting heart-beat rate, and instantaneous blood flow velocities from 0.7 to 20 mm/s were measured.
描述了使用自适应光学部分共焦多线检眼镜(AO-pcMLO)进行高速、相衬视网膜和血流成像的方法。它通过使用单个二维相机和数字微镜器件(DMD),实现了利用各种方向的多线照明同时进行共焦和相衬成像。对垂直和水平方向的线照明都进行了测试,用于光感受器和血管成像。相衬成像能够更好地显示视网膜结构,如视锥细胞内段、血管壁和红细胞,成像帧率高达500 Hz。使用毛细管记波图测量直径小于40μm的小血管的血流速度,对小动脉或小静脉则测量后续图像之间的互相关。在正常静息心率下观察到与心脏相关的搏动模式,测量到的瞬时血流速度为0.7至20 mm/s。