Med Eng Phys. 2024 Mar;125:104125. doi: 10.1016/j.medengphy.2024.104125. Epub 2024 Feb 20.
Proprioceptive function assessment is crucial in clinical practice for patients with chronic non-specific neck pain (CNNP) as it is a major issue affecting their condition.
To verify the reliability and validity of baiobit sensor in measuring the neck proprioceptive function of CNNP patients.
Fifty-three CNNP patients were recruited (36 females, 17 males; age range 21-60 years) and were assessed for cervical joint position error by two blinded raters using the Baiobit sensor and laser pointer devices. The second measurement was conducted by the same rater 48 h later. Intra and inter-rater reliability of the Baiobit sensor was evaluated using the intra-class correlation coefficient (ICC), while the validity of the Baiobit sensor was established using the Spearman correlation coefficient.
The Baiobit sensor demonstrated moderate to excellent intra-rater reliability in flexion, extension, left lateral flexion, right lateral flexion, and right rotation (ICCs=0.71∼0.85, 95 %CIs: 0.50∼0.91), left-rotation shows poor to good intra-rater reliability (ICC=0.56, 95 %CI: 0.25∼0.75). The Baiobit sensor also demonstrated moderate to excellent inter-rater reliability in flexion, extension, left lateral flexion, right lateral flexion, and right rotation (ICCs=0.80∼0.88, 95 %CIs: 0.65∼0.91), left-rotation shows poor to good intra-rater reliability (ICC=0.59, 95 %CI: 0.29∼0.76). Validity analysis showed that the Baiobit sensor had a range of low to high validity (r = 0.46∼0.88) for measuring cervical proprioception function, with lower validity observed in the left flexion direction. The Baiobit showed good absolute reliability with low SEM and MDC values (0.35°∼2.42°).
The new device could be used as an alternative tool to evaluate neck proprioception.
本体感觉功能评估对于慢性非特异性颈痛(CNNP)患者的临床实践至关重要,因为这是影响其病情的主要问题。
验证 baiobit 传感器测量 CNNP 患者颈部本体感觉功能的可靠性和有效性。
招募了 53 名 CNNP 患者(36 名女性,17 名男性;年龄 21-60 岁),由两名盲法评估者使用 baiobit 传感器和激光笔设备评估颈椎关节位置误差。第二次测量由同一位评估者在 48 小时后进行。采用组内相关系数(ICC)评估 baiobit 传感器的内部和内部可靠性,采用 Spearman 相关系数确定 baiobit 传感器的有效性。
baiobit 传感器在屈伸、左、右侧屈和右旋转方向具有中度至高度的内部可靠性(ICC=0.71∼0.85,95%CI:0.50∼0.91),左侧旋转显示出从差到好的内部可靠性(ICC=0.56,95%CI:0.25∼0.75)。baiobit 传感器在屈伸、左、右侧屈和右旋转方向也具有中度至高度的内部可靠性(ICC=0.80∼0.88,95%CI:0.65∼0.91),左侧旋转显示出从差到好的内部可靠性(ICC=0.59,95%CI:0.29∼0.76)。有效性分析表明,baiobit 传感器在测量颈椎本体感觉功能方面具有从低到高的有效性(r=0.46∼0.88),左侧屈曲方向的有效性较低。baiobit 具有良好的绝对可靠性,SEM 和 MDC 值较低(0.35°∼2.42°)。