Ohyama Yohei, Minoda Yukihide, Masuda Sho, Sugama Ryo, Ohta Yoichi, Nakamura Hiroaki
Department of Orthopaedic Surgery, Osaka Metropolitan University Graduate School of Medicine, 1-4-3 Asahi-Machi Abeno-Ku Osaka-City, Osaka, 545-8585, Japan.
Eur J Orthop Surg Traumatol. 2024 May;34(4):2041-2047. doi: 10.1007/s00590-024-03891-0. Epub 2024 Mar 22.
This study aimed to compare bone mineral density (BMD) changes around the femoral component after total hip arthroplasty (THA) in a fully hydroxyapatite-coated stem (CORAIL) and in a tapered-wedge stem (Taperloc complete) and identify predictors of BMD changes.
This retrospective study compared 43 hips in the CORAIL group and 40 hips in the Taperloc group. The relative changes in BMD at 2 years after THA measured using dual-energy X-ray absorptiometry and the three-dimensional quantified contact states of the stem with the femoral cortical bone were assessed. Predictors of the relative change in the BMD around the proximal part of the stem were examined using multiple regression analysis.
The decrease in BMD in Gruen zone 7 was significantly less in the CORAIL group than in the Taperloc group (P = 0.02). In the CORAIL group, the contact area in any zone was not a significant predictor of the relative changes in BMD. The contact area between the Taperloc stem and the femoral cortical bone in zones 2 and 6 was a positive predictor of the relative changes in BMD in zones 1 (P = 0.02 and P = 0.04, respectively) and 2 (P = 0.008 and P = 0.004, respectively).
The CORAIL stem suppressed the postoperative BMD loss around the stem, irrespective of the contact state. The Taperloc complete stem required contact with the proximal femoral metaphysis to suppress the postoperative BMD loss around the stem.
本研究旨在比较全羟基磷灰石涂层柄(CORAIL)和锥形楔形柄(Taperloc complete)在全髋关节置换术(THA)后股骨部件周围的骨密度(BMD)变化,并确定BMD变化的预测因素。
这项回顾性研究比较了CORAIL组的43例髋关节和Taperloc组的40例髋关节。评估了使用双能X线吸收法测量的THA术后2年时BMD的相对变化以及柄与股骨皮质骨的三维定量接触状态。使用多元回归分析检查柄近端周围BMD相对变化的预测因素。
CORAIL组Gruen 7区的BMD下降明显少于Taperloc组(P = 0.02)。在CORAIL组中,任何区域的接触面积都不是BMD相对变化的显著预测因素。Taperloc柄与股骨皮质骨在2区和6区的接触面积是1区(分别为P = 0.02和P = 0.04)和2区(分别为P = 0.008和P = 0.004)BMD相对变化的正向预测因素。
无论接触状态如何,CORAIL柄均能抑制柄周围术后BMD的丢失。Taperloc complete柄需要与股骨近端干骺端接触以抑制柄周围术后BMD的丢失。