Suppr超能文献

微圈套器取出术作为游离编织型血管内桥接装置的补救技术:病例系列

Microsnare retrieval as a bail-out technique of Detached Woven EndoBridge device: Illustrative series.

作者信息

Chew Han Seng, Altibi Mohammad, Al-Ali Samer, Butler Benjamin, Butt Waleed, Chavda Swarupsinh, Lamin Saleh

机构信息

Interventional Neuroradiology Department, Queen Elizabeth Hospital Birmingham, Birmingham, UK.

出版信息

Interv Neuroradiol. 2024 Apr 5:15910199241242170. doi: 10.1177/15910199241242170.

Abstract

BACKGROUND

The Woven EndoBridge (WEB) device (MicroVention, Tustin, CA, USA) has an excellent safety profile. While major complications such as device malposition and migration are rare, they can have serious consequences if not addressed promptly. Our case series describes the safety and efficacy of Amplatz goose neck microsnare device (Medtronic in Irvine, CA, USA) in endovascular retrieval of a detached WEB device.

METHODS

We retrospectively reviewed six consecutive patients who underwent endovascular WEB retrieval using Amplatz microsnare device between March 2012 and December 2022.

RESULTS

All six WEB devices were successfully retrieved either directly from the aneurysm sac due to device malpositioning or from a distal branch following device migration. None of the patients experienced intra-operative aneurysm perforation, arterial dissection, or vasospasm attributable to the process of WEB extraction. Five out of six patients (83.3%) had a good functional outcome (mRS 0-1) upon discharge from the hospital and at 24 months.

CONCLUSION

Our experience suggests that detached WEB devices can be safely retrieved using an Amplatz microsnare. Apart from addressing device migration, direct removal of an undersized or malpositioned WEB from the aneurysm sac appears to be a safe option that can be considered when all other rescue techniques have been exhausted.

摘要

背景

编织型血管内桥接器(WEB)装置(美国加利福尼亚州图斯廷市的MicroVention公司生产)具有出色的安全性。虽然诸如装置位置不当和移位等主要并发症很少见,但如果不及时处理,可能会产生严重后果。我们的病例系列描述了用于血管内取出分离的WEB装置的安普拉斯鹅颈微圈套器(美国加利福尼亚州欧文市的美敦力公司生产)的安全性和有效性。

方法

我们回顾性分析了2012年3月至2022年12月期间连续6例使用安普拉斯微圈套器进行血管内WEB取出术的患者。

结果

由于装置位置不当,所有6个WEB装置均直接从动脉瘤腔内成功取出,或在装置移位后从远端分支取出。所有患者在WEB取出过程中均未发生术中动脉瘤穿孔、动脉夹层或血管痉挛。6例患者中有5例(83.3%)在出院时及24个月时功能预后良好(改良Rankin量表评分0 - 1分)。

结论

我们的经验表明,使用安普拉斯微圈套器可以安全地取出分离的WEB装置。除了解决装置移位问题外,当所有其他抢救技术均已用尽时,直接从动脉瘤腔内取出尺寸过小或位置不当的WEB似乎是一种可以考虑的安全选择。

相似文献

6
Snare salvage technique for deformed WEB device after deployment.圈套回收技术在部署后变形 WEB 装置。
J Neurointerv Surg. 2021 Mar;13(3):294. doi: 10.1136/neurintsurg-2020-016696. Epub 2020 Sep 8.

本文引用的文献

7
Snare salvage technique for deformed WEB device after deployment.圈套回收技术在部署后变形 WEB 装置。
J Neurointerv Surg. 2021 Mar;13(3):294. doi: 10.1136/neurintsurg-2020-016696. Epub 2020 Sep 8.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验