Med Phys. 2024 May;51(5):3796-3805. doi: 10.1002/mp.17060. Epub 2024 Apr 8.
The Relative Biological Effectiveness (RBE) of kilovoltage photon beams has been previously investigated in vitro and in silico using analytical methods. The estimated values range from 1.03 to 1.82 depending on the methodology and beam energies examined.
The focus of this work was to independently estimate RBE values for a range of clinically used kilovoltage beams (70-200 kVp) while investigating the suitability of using TOPAS-nBio for this task.
Previously validated spectra of clinical beams were used to generate secondary electron spectra at several depths in a water tank phantom via TOPAS Monte Carlo (MC) simulations. Cell geometry was irradiated with the secondary electrons in TOPAS-nBio MC simulations. The deposited dose and the calculated number of DNA strand breaks were used to estimate RBE values.
Monoenergetic secondary electron simulations revealed the highest direct and indirect double strand break yield at approximately 20 keV. The average RBE value for the kilovoltage beams was calculated to be 1.14.
TOPAS-nBio was successfully used to estimate the RBE values for a range of clinical radiotherapy beams. The calculated value was in agreement with previous estimates, providing confidence in its clinical use in the future.
先前已经使用分析方法在体外和计算机模拟中研究了千伏光子束的相对生物效应(RBE)。根据所检查的方法和射束能量,估计值的范围从 1.03 到 1.82 不等。
本工作的重点是独立估算一系列临床使用的千伏射束(70-200 kVp)的 RBE 值,同时研究 TOPAS-nBio 在此任务中的适用性。
使用先前经过验证的临床射束谱,通过 TOPAS 蒙特卡罗(MC)模拟在水箱体模中的几个深度处生成二次电子谱。在 TOPAS-nBio MC 模拟中,用二次电子照射细胞几何形状。使用沉积剂量和计算的 DNA 链断裂数来估算 RBE 值。
单能二次电子模拟显示在大约 20 keV 时具有最高的直接和间接双链断裂产额。估算的千伏射线束的平均 RBE 值为 1.14。
TOPAS-nBio 成功用于估算一系列临床放射治疗射束的 RBE 值。计算值与先前的估计值一致,为其未来在临床中的应用提供了信心。