Department of Microbiology, Maulana Azad Medical College & Associated Lok Nayak Hospital, Bahadur Shah Zafar Marg, New Delhi, 110002, India; State Level -Viral Research and Diagnostic Laboratory (VRDL), Department of Microbiology, Maulana Azad Medical College & Associated Lok Nayak Hospital, Bahadur Shah Zafar Marg, New Delhi, 110002, India.
Department of Microbiology, Maulana Azad Medical College & Associated Lok Nayak Hospital, Bahadur Shah Zafar Marg, New Delhi, 110002, India.
Indian J Med Microbiol. 2024 May-Jun;49:100577. doi: 10.1016/j.ijmmb.2024.100577. Epub 2024 Apr 13.
Indian Council of Medical Research (ICMR) initiated an Inter-Laboratory Quality Control testing (ILQC) program for Severe Acute Respiratory Syndrome Coronavirus 2 (SARS-CoV-2) testing. Under this program, SARS-CoV-2 testing laboratories across the country submit specimens to the assigned State Quality Control (SQCs) laboratories for ILQC testing. This study aimed to investigate the performance of public and private SARS-CoV-2 testing laboratories in Delhi and highlights the country's effort in ramping up testing facility with close monitoring of the quality of Covid-19 testing results.
In the present study, two-years of SARS-CoV-2 testing data is included. During July 2020 through February 2022, a total of 1791 anonymised specimens were received from 56 public and private laboratories. These specimens were processed by reverse transcriptase - polymerase chain reaction (RT-PCR) tests as per National Institute of Virology (NIV) protocol and the results were uploaded on the ICMR quality control/quality assurance (QC/QA) portal without directly conveying the results to respective participating laboratories. This portal generated a final report stating concordance and intimate results to individual laboratories.
Among the 1791 specimens, 25 were rejected and the remaining 1766 were tested. Among these specimens 1691 (95.75%) revealed concordance, and 75 (4.24%) were discordant. A total of 29 laboratories had 100% concordance, 21 laboratories had over 90% concordance and six laboratories had over 80% concordance.
The study demonstrates that the establishment of an inter-laboratory comparison program for SARS-CoV-2 testing helped in monitoring quality of SARS-CoV-2 testing in the country.
印度医学研究理事会(ICMR)启动了一项针对严重急性呼吸系统综合征冠状病毒 2(SARS-CoV-2)检测的实验室间质量控制测试(ILQC)计划。根据该计划,全国各地的 SARS-CoV-2 检测实验室将标本提交给指定的州质量控制(SQC)实验室进行 ILQC 测试。本研究旨在调查德里的公立和私立 SARS-CoV-2 检测实验室的性能,并强调了该国在提高检测设施的同时密切监测 COVID-19 检测结果质量的努力。
本研究包括两年的 SARS-CoV-2 检测数据。2020 年 7 月至 2022 年 2 月期间,共收到来自 56 家公立和私立实验室的 1791 份匿名标本。这些标本按照国家病毒学研究所(NIV)的方案通过逆转录-聚合酶链反应(RT-PCR)检测进行处理,结果在 ICMR 质量控制/质量保证(QC/QA)门户上上传,而不直接向各自参与的实验室传达结果。该门户生成了一份最终报告,向各个实验室说明一致性和具体结果。
在 1791 个标本中,有 25 个被拒收,其余 1766 个进行了检测。在这些标本中,有 1691 个(95.75%)显示一致性,有 75 个(4.24%)不一致。共有 29 家实验室的一致性达到 100%,21 家实验室的一致性超过 90%,6 家实验室的一致性超过 80%。
研究表明,建立 SARS-CoV-2 检测实验室间比较计划有助于监测该国 SARS-CoV-2 检测的质量。