Suppr超能文献

机械通气时的压缩容积:呼吸机与管路系统的比较

Compression volume during mechanical ventilation: comparison of ventilators and tubing circuits.

作者信息

Bartel L P, Bazik J R, Powner D J

出版信息

Crit Care Med. 1985 Oct;13(10):851-4. doi: 10.1097/00003246-198510000-00016.

Abstract

Four ventilators (Puritan-Bennett MA-1 and MA-2, Emerson, and Bear I) and four commercially available disposable and nondisposable tubing circuits (Bennett nondisposable, Becton-Dickinson, Inspiron, and Life-line) were tested on a lung analog for differences in inspiratory-circuit compression volume. The compression ratio (Rc), equal to the gas volume compressed per cm H2O peak airway pressure, was calculated for each combination of ventilator and circuit at each of four compliance settings (0.15, 0.10, 0.05, 0.01 L/cm H2O) on the analog. Rc values ranged from 0.3 to 4.5 ml/cm H2O at the highest and lowest compliance settings, respectively, accounting for a reduction in delivered tidal volume of up to 20%. The Emerson ventilator with all tubing systems and the Bennett nondisposable circuit with each ventilator demonstrated slightly smaller compression volumes. Application of an inspiratory pause on the Bear I ventilator did not affect its compression characteristics. The clinical importance of compression volume and data from other ventilation systems are reviewed.

摘要

对四台呼吸机(普里坦-贝内特MA-1和MA-2、爱默生和熊牌I型)以及四种市售的一次性和非一次性管路回路(贝内特非一次性、贝克顿-迪金森、英斯派尔和生命线)在肺模拟装置上进行了吸气回路压缩容积差异测试。针对模拟装置上四种顺应性设置(0.15、0.10、0.05、0.01 L/cm H₂O)中的每一种,计算了呼吸机与回路的每种组合的压缩比(Rc),其等于每厘米水柱峰值气道压力下压缩的气体容积。在最高和最低顺应性设置下,Rc值分别为0.3至4.5 ml/cm H₂O,这导致输送潮气量减少高达20%。配备所有管路系统的爱默生呼吸机以及与每种呼吸机搭配的贝内特非一次性回路的压缩容积略小。在熊牌I型呼吸机上应用吸气暂停并不影响其压缩特性。本文回顾了压缩容积的临床重要性以及来自其他通气系统的数据。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验