Truzzi Stefano, Manns Fabrice, Parel Jean-Marie, Ruggeri Marco
Ophthalmic Biophysics Center, Bascom Palmer Eye Institute, University of Miami Miller School of Medicine, Miami, FL, USA.
Department of Engineering "Enzo Ferrari", University of Modena and Reggio Emilia, Modena, Italy.
Biomed Opt Express. 2024 Mar 29;15(4):2681-2696. doi: 10.1364/BOE.520735. eCollection 2024 Apr 1.
We present proof of concept for a handheld contact-type system capable of simultaneous optical coherence tomography (OCT) imaging of the retina and wide-field digital fundus color photography. The study focuses on demonstrating the feasibility of the proposed approach, particularly for eventual use in pediatric patients during examination under anesthesia in the operating room and in the neonatal intensive care unit. Direct contact of the probe with the cornea allows the photographer to maintain a stable position during imaging, reducing motion artifacts in the OCT images. Additionally, it simplifies the alignment process and increases the field of view of the optics. By integrating OCT and fundus imaging into a single device, the proposed compact modular design eliminates the need for separate, space-consuming systems dedicated to each imaging modality.
我们展示了一种手持式接触式系统的概念验证,该系统能够同时对视网膜进行光学相干断层扫描(OCT)成像和广角数字眼底彩色摄影。该研究重点在于证明所提出方法的可行性,特别是最终用于手术室麻醉检查期间的儿科患者以及新生儿重症监护病房。探头与角膜的直接接触使摄影师在成像过程中能够保持稳定位置,减少OCT图像中的运动伪影。此外,它简化了对准过程并增加了光学系统的视野。通过将OCT和眼底成像集成到单个设备中,所提出的紧凑模块化设计消除了对每种成像方式专用的单独、占空间的系统的需求。