连续葡萄糖监测无校准组织间液传感器与血糖仪毛细血管血糖监测的血糖读数差异:两例分析

Differences in glucose readings between the continuous glucose monitoring calibration free interstitial sensors versus capillary blood glucose monitoring by glucometer: An analysis of two cases.

作者信息

Quah Phaik Ling, Chai Sally Mun Hua, Tan Kok Hian

机构信息

Division of Obstetrics & Gynaecology, KK Women's and Children's Hospital, Singapore.

Duke-NUS Medical School, National University of Singapore, Singapore.

出版信息

Metabol Open. 2024 Apr 18;22:100282. doi: 10.1016/j.metop.2024.100282. eCollection 2024 Jun.

Abstract

AIM

To assess the differences in glucose readings between the continuous glucose monitoring calibration-free interstitial sensors versus capillary blood glucose monitoring by glucometer.

STUDY DESIGN

Two healthy non-pregnant volunteers participated in the study, and wore simultaneously both the calibration-free Freestyle Libre and the Dexcom G6 sensor. Glucose values were recorded before and after meals during breakfast, lunch, and dinner on three separate days by either scanning the Freestyle Libre CGM sensor with a smartphone, or obtaining glucose readings real-time through the Dexcom G6 CLARITY mobile application. Blood glucose values were recorded using the Accu-Chek Active glucose meter. The Wilcoxon signed-rank test was used for paired non-parametric data to compare glucose readings between groups.

RESULTS

The average glucose values obtained from the Dexcom G6 CGM consistently registered higher (6.54 ± 0.80 mmol/L) and those from the Freestyle Libre (5.49 ± 0.65 mmol/L) consistently lower, from the glucometer (6.17 ± 0.55 mmol/L), with p-value <0.05 between groups. In the three-way comparison, the Dexcom G6 CGM sensor yielded the highest values, followed by the glucose meter, and finally the Freestyle Libre CGM sensor.

CONCLUSION

Both CGM systems exhibited discrepancies from blood glucose (BG) measurements, and variations were observed among the different CGM systems themselves.

摘要

目的

评估连续葡萄糖监测免校准组织间液传感器与血糖仪毛细血管血糖监测之间的血糖读数差异。

研究设计

两名健康非孕志愿者参与研究,同时佩戴免校准的Freestyle Libre和德康G6传感器。在三个不同日期的早餐、午餐和晚餐期间,通过用智能手机扫描Freestyle Libre连续葡萄糖监测(CGM)传感器,或通过德康G6 CLARITY移动应用程序实时获取血糖读数,记录饭前饭后的葡萄糖值。使用拜安时活力血糖仪记录血糖值。采用Wilcoxon符号秩检验对配对非参数数据进行分析,以比较各组之间的血糖读数。

结果

德康G6 CGM获得的平均葡萄糖值始终较高(6.54±0.80毫摩尔/升),Freestyle Libre的平均葡萄糖值(5.49±0.65毫摩尔/升)始终低于血糖仪(6.17±0.55毫摩尔/升),组间p值<0.05。在三方比较中,德康G6 CGM传感器的值最高,其次是血糖仪,最后是Freestyle Libre CGM传感器。

结论

两种CGM系统的血糖测量结果均存在差异,且不同CGM系统之间也存在差异。

相似文献

6
Inaccurate Glucose Sensor Values After Hydroxyurea Administration.羟脲给药后葡萄糖传感器值不准确。
Diabetes Technol Ther. 2021 Jun;23(6):443-451. doi: 10.1089/dia.2020.0490. Epub 2020 Dec 30.

本文引用的文献

4
Freestyle Libre Glucose Monitoring System.免调码型葡萄糖监测系统
Clin Diabetes. 2018 Apr;36(2):203-204. doi: 10.2337/cd17-0130.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验