用于评估无标记体表引导放射治疗中摆位精度的体模研制。

Development of a phantom for assessing the precision of setup in skin mark-less surface-guided radiotherapy.

机构信息

Department of Radiology, University of Yamanashi, Yamanashi, Japan.

Kasugai General Rehabilitation Hospital, Yamanashi, Japan.

出版信息

J Appl Clin Med Phys. 2024 Aug;25(8):e14381. doi: 10.1002/acm2.14381. Epub 2024 May 2.

Abstract

BACKGROUND

Surface-guided radiotherapy (SGRT) is adopted by several institutions; however, reports on the phantoms used to assess the precision of the SGRT setup are limited.

PURPOSE

The purpose of this study was to develop a phantom to verify the accuracy of the irradiation position during skin mark-less SGRT.

METHODS

An acrylonitrile butadiene styrene (ABS) plastic cube phantom with a diameter of 150 mm on each side containing a dummy target of 15 mm and two types of body surface-shaped phantoms (breast/face shape) that could be attached to the cube phantom were fabricated. Films can be inserted on four sides of the cubic phantom (left, right, anterior and posterior), and the center of radiation can be calculated by irradiating the dummy target with orthogonal MV beams. Three types of SGRT using a VOXELAN-HEV600M (Electronics Research&Development Corporation, Okayama, Japan) were evaluated using this phantom: (i) SGRT-a SGRT set-up based solely on a computed tomography (CT)-reference image. (ii) SGRT + CBCT-a method where cone beam computed tomography (CBCT) matching was performed after SGRT. (iii) SGRT-a resetup technique using a scan reference image obtained after completing the (ii) step.

RESULTS

Both the breast and face phantoms were recognized in the SGRT system without problems. SGRT ensure precision within 1 mm/1° for breast and face verification, respectively. All SGRT methods showed comparable rotational accuracies with no significant disparities.

CONCLUSIONS

The developed phantom was useful for verifying the accuracy of skin mark-less SGRT position matching. The SGRT demonstrated the feasibility of achieving skin-mark less SGRT with high accuracy, with deviations of less than 1 mm. Additional research is necessary to evaluate the suitability of the developed phantoms for use in various facilities and systems. This phantom could be used for postal surveys in the future.

摘要

背景

一些机构采用表面引导放疗(SGRT);然而,用于评估 SGRT 设置精度的体模的报告有限。

目的

本研究的目的是开发一种体模,以验证无皮肤标记 SGRT 中照射位置的准确性。

方法

制作了一个直径为 150mm 的丙烯腈丁二烯苯乙烯(ABS)塑料立方休模体,每边包含一个 15mm 的模拟靶和两个可附着在立方休模体上的体表面形状模体(乳房/面部形状)。立方休模体的四个侧面(左侧、右侧、前侧和后侧)可插入胶片,通过对模拟靶进行正交 MV 射线照射可以计算出放射中心。使用 VOXELAN-HEV600M(日本电子研发公司)对三种类型的 SGRT 进行了评估:(i)SGRT-a 仅基于计算机断层扫描(CT)参考图像的 SGRT 设置。(ii)SGRT+CBCT-a 在 SGRT 后进行锥形束 CT(CBCT)匹配的方法。(iii)使用完成(ii)步骤后获得的扫描参考图像进行的 SGRT-a 重新设置技术。

结果

乳房和面部模体在 SGRT 系统中均能无问题地识别。SGRT 分别确保了乳房和面部验证的精度在 1mm/1°以内。所有 SGRT 方法的旋转精度相当,没有显著差异。

结论

所开发的体模可用于验证无皮肤标记 SGRT 位置匹配的准确性。SGRT 证明了实现高精度无皮肤标记 SGRT 的可行性,偏差小于 1mm。需要进一步研究来评估所开发的体模在各种设施和系统中的适用性。该体模将来可用于邮寄调查。