John Moran Eye Center, University of Utah, Salt Lake City, Utah, United States.
Clinica de Olhos De Francesco, Fortaleza, Brazil.
Invest Ophthalmol Vis Sci. 2024 May 1;65(5):13. doi: 10.1167/iovs.65.5.13.
The purpose of this study was to measure intraocular pressure (IOP) elevation while applying standard gonioscopy, selective laser trabeculoplasty (SLT), and laser iridotomy procedural lenses.
Twelve cadaver eyes were mounted to a custom apparatus and cannulated with a pressure transducer which measured IOP. The apparatus was mounted to a load cell which measured the force on the eye. Six ophthalmologists performed simulated gonioscopy (Sussman 4 mirror lens), SLT (Latina lens), and laser iridotomy (Abraham lens) while a computer recorded IOP (mm Hg) and force (grams). The main outcome measures were IOP and force applied to the eye globe during ophthalmic diagnostics and procedures.
The average IOP's during gonioscopy, SLT, and laser iridotomy were 43.2 ± 16.9 mm Hg, 39.8 ± 9.9 mm Hg, and 42.7 ± 12.6 mm Hg, respectively. The mean force on the eye for the Sussman, Latina, and Abraham lens was 40.3 ± 26.4 grams, 66.7 ± 29.8 grams, and 65.5 ± 35.9 grams, respectively. The average force applied to the eye by the Sussman lens was significantly lower than both the Latina lens (P = 0.0008) and the Abraham lens (P = 0.001). During gonioscopy indentation, IOP elevated on average to 80.5 ± 22.6 mm Hg. During simulated laser iridotomy tamponade, IOP elevated on average to 82.3 ± 27.2 mm Hg.
In cadaver eyes, the use of standard ophthalmic procedural lenses elevated IOP by approximately 20 mm Hg above baseline.
本研究旨在测量标准房角镜检查、选择性激光小梁成形术(SLT)和激光虹膜切开术程序镜片应用时的眼压(IOP)升高情况。
将 12 只尸体眼球安装到一个定制的装置上,并通过压力传感器进行套管穿刺,该传感器用于测量 IOP。该装置安装在一个称重传感器上,用于测量眼球上的力。六名眼科医生进行模拟房角镜检查(Sussman 4 镜)、SLT(Latina 镜片)和激光虹膜切开术(Abraham 镜片),同时计算机记录 IOP(mmHg)和施加在眼球上的力(克)。主要观察指标是眼科诊断和手术过程中眼压(mmHg)和眼球受力(克)。
房角镜检查、SLT 和激光虹膜切开术的平均 IOP 分别为 43.2 ± 16.9mmHg、39.8 ± 9.9mmHg 和 42.7 ± 12.6mmHg。Sussman、Latina 和 Abraham 镜片对眼球的平均作用力分别为 40.3 ± 26.4 克、66.7 ± 29.8 克和 65.5 ± 35.9 克。Sussman 镜片施加在眼球上的平均力明显低于 Latina 镜片(P=0.0008)和 Abraham 镜片(P=0.001)。在房角镜检查时,眼压平均升高到 80.5 ± 22.6mmHg。在模拟激光虹膜切开术压迫时,眼压平均升高到 82.3 ± 27.2mmHg。
在尸体眼球中,标准眼科手术镜片的使用使 IOP 升高约 20mmHg 以上。