Nobiling R, Bührle C P, Taugner R
Histochemistry. 1985;82(1):45-9. doi: 10.1007/BF00502090.
Freeze-fracture autoradiography is accepted as an adequate technique for localization studies of soluble substances at the electron microscopical level. The method, however, involves many critical preparation steps, among them a protective carbon coating of the developed nuclear emulsion adhering to the replica. We demonstrate here that this additional carbon coating may be omitted. This simplification leads to a significant improvement of the sample yield as compared with the previously described procedures.
冷冻断裂放射自显影术被认为是在电子显微镜水平上对可溶性物质进行定位研究的一种适当技术。然而,该方法涉及许多关键的制备步骤,其中包括在附着于复制品的显影核乳剂上进行保护性碳涂层处理。我们在此证明,这种额外的碳涂层可以省略。与先前描述的程序相比,这种简化导致样品产量显著提高。