Wandtke J C, Plewes D B
AJR Am J Roentgenol. 1985 Jun;144(6):1171-81. doi: 10.2214/ajr.144.6.1171.
Exposure equalization radiography systems with scanning slit and raster geometries were constructed and tested with 75 patients. The scanning equalization radiography (SER) technique uses a detector, placed behind the patient, connected in a feedback loop to a microprocessor-controlled x-ray source. The detector monitors the transmitted radiation, and in turn the x-ray output is varied to equalize the radiographic film density over the entire image. The clinical evaluation of these systems included 25 posteroanterior (PA) chest radiographs by an SER slit-geometry system (5.0-sec scan time), 25 PA chest radiographs by an SER raster-geometry system with an 8.8-sec scan time, and 25 PA chest radiographs by an SER raster-geometry system with a 4.7-sec scan time. These SER radiographs were compared to conventional radiographs of the same patients by two radiologists. The observers noted that the SER slit radiographs had seriously overexposed areas in 80% of instances, and that any potential gains from this system were offset by the overexposure problems. The radiographs obtained by the SER raster technique with a 4.7-sec scan time showed more uniform and adequate exposure in 80% of instances and better visualization of normal anatomic detail in the lung (52%) and mediastinum (84%) than conventional radiographs. The radiographs obtained by the SER raster technique with an 8.8-sec scan time showed fewer peripheral lung markings in 15 of 25 cases, presumably due to motion. In all other respects, the images were similar in quality to the SER raster 4.7-sec radiographs.
构建了具有扫描狭缝和光栅几何结构的曝光均衡射线照相系统,并对75名患者进行了测试。扫描均衡射线照相(SER)技术使用一个置于患者身后的探测器,该探测器通过反馈回路与微处理器控制的X射线源相连。探测器监测透射辐射,进而改变X射线输出,以使整个图像上的射线照相胶片密度均匀。这些系统的临床评估包括由一个SER狭缝几何结构系统(扫描时间为5.0秒)拍摄25张后前位(PA)胸部射线照片、由一个扫描时间为8.8秒的SER光栅几何结构系统拍摄25张PA胸部射线照片以及由一个扫描时间为4.7秒的SER光栅几何结构系统拍摄25张PA胸部射线照片。两位放射科医生将这些SER射线照片与同一患者的传统射线照片进行了比较。观察者指出,SER狭缝射线照片在80%的情况下存在严重过度曝光区域,并且该系统的任何潜在优势都被过度曝光问题抵消了。扫描时间为4.7秒的SER光栅技术所获得的射线照片在80%的情况下显示出更均匀和适当的曝光,并且与传统射线照片相比,在肺部(52%)和纵隔(84%)中正常解剖细节的显示更好。扫描时间为8.8秒的SER光栅技术所获得的射线照片在25例中有15例显示外周肺纹理较少,推测是由于运动所致。在所有其他方面,这些图像的质量与扫描时间为4.7秒的SER光栅射线照片相似。