Senda M, Tamaki N, Yonekura Y, Tanada S, Murata K, Hayashi N, Fujita T, Saji H, Konishi J, Torizuka K
J Comput Assist Tomogr. 1985 Sep-Oct;9(5):940-6. doi: 10.1097/00004728-198509000-00018.
A whole-body, multislice positron emission tomograph, Positologica III, has been constructed based on a sampling method "positology." The scanner consists of four continuously rotating detector rings with 192 bismuth germanate (BGO) crystals (12 X 24 X 24 mm) arranged at irregular intervals along each ring, providing seven tomographic images at 16 mm intervals simultaneously, for a total axial coverage of 12 cm. The patient aperture is 54 cm in diameter, and the field of view is 40 cm in diameter and 12 cm in depth. Newly developed dual photomultiplier tubes are used with 12 mm wide BGO crystals to provide a packing ratio of 0.894. The spatial resolution at the center of the field is 7.6 mm full width at half maximum, and 2.5 mm hot spots were delineated in the Derenzo phantom. The sensitivity for a 20 cm diameter cylindrical phantom is 34.2 and 52.2 kcps/muCi/ml for true planes and cross planes, respectively. In clinical studies the posterior papillary muscle of the heart was visualized in the myocardial scan, and the lung behind the pulmonary artery was also visualized in the pulmonary ventilation scan. These results suggest that our machine has sufficient resolution for most of the clinical studies of the body.
基于一种“正电子断层学”采样方法构建了一台全身多层正电子发射断层扫描仪,即Positologica III。该扫描仪由四个连续旋转的探测器环组成,每个环上以不规则间隔排列着192块锗酸铋(BGO)晶体(12×24×24毫米),可同时以16毫米的间隔提供七张断层图像,轴向总覆盖范围为12厘米。患者孔径直径为54厘米,视野直径为40厘米,深度为12厘米。新开发的双光电倍增管与12毫米宽的BGO晶体配合使用,填充率为0.894。视野中心的空间分辨率为半高宽7.6毫米,在德伦佐体模中可分辨出2.5毫米的热点。对于直径20厘米的圆柱形体模,真平面和交叉平面的灵敏度分别为34.2和52.2千计数/微居里/毫升。在临床研究中,心脏的后乳头肌在心肌扫描中可见,肺动脉后方的肺在肺通气扫描中也可见。这些结果表明,我们的机器对于身体的大多数临床研究具有足够的分辨率。