Miyasaka Yuya, Kanai Takayuki, Souda Hikaru, Yamazawa Yoshifumi, Lee Sung Hyun, Chai Hongbo, Sato Hiraku, Iwai Takeo
Department of Heavy Particle Medical Science, Yamagata University Graduate School of Medical Science, Yamagata, Japan.
Department of Radiation Oncology, Tokyo Women's Medical University, Shinjuku, Tokyo, Japan.
Int J Part Ther. 2024 Apr 25;11:100011. doi: 10.1016/j.ijpt.2024.100011. eCollection 2024 Mar.
We performed computed tomography (CT)-stopping power ratio (SPR) calibration in a carbon-ion therapy facility and evaluated SPR estimation accuracy.
A polybinary tissue model method was used for the calibration of CT numbers and SPR. As a verification by dose calculation, we created a virtual phantom to which the CT-SPR calibration table was applied. Then, SPR was calculated from the change in the range of the treatment planning beam when changing to 19 different CT numbers, and the accuracy of the treatment planning system (TPS) calculation of SPR values from the CT-SPR calibration table was validated. As a verification by measurement, 5 materials (water, milk, olive oil, ethanol, 40% KHPO) were placed in a container, and the SPR was obtained by measurement from the change in the range of the beam that passed through the materials.
The results of the dose calculations of the TPS showed that the results agreed within 1% for the lower CT numbers up to 1000 HU, but there was a difference of 3.0% in the higher CT number volume. The difference between the SPR calculated by TPS and the SPR caused by the difference in the energy of the incident particles agreed within 0.51%. The accuracy of SPR estimation was measured, and the error was within 2% for all materials tested.
These results indicate that the SPR estimation errors are within the range of errors that can be expected in particle therapy. From commissioning and verification results, the CT-SPR calibration table obtained during this commissioning process is clinically applicable.
我们在碳离子治疗设备中进行了计算机断层扫描(CT)阻止本领比(SPR)校准,并评估了SPR估计精度。
采用多元组织模型方法对CT值和SPR进行校准。作为剂量计算验证,我们创建了一个应用CT-SPR校准表的虚拟体模。然后,当改变为19种不同CT值时,根据治疗计划射束射程的变化计算SPR,并验证了治疗计划系统(TPS)根据CT-SPR校准表计算SPR值的准确性。作为测量验证,将5种材料(水、牛奶、橄榄油、乙醇、40%KHPO)置于容器中,通过测量穿过这些材料的射束射程变化来获得SPR。
TPS的剂量计算结果表明,对于CT值低于1000HU的情况,结果在1%以内相符,但在较高CT值区域存在3.0%的差异。TPS计算的SPR与由入射粒子能量差异引起的SPR之间的差异在0.51%以内相符。测量了SPR估计的准确性,所有测试材料的误差均在2%以内。
这些结果表明,SPR估计误差在粒子治疗预期的误差范围内。从调试和验证结果来看,本次调试过程中获得的CT-SPR校准表在临床上是适用的。