Gundberg C M, Wilson P S, Gallop P M, Parfitt A M
Clin Chem. 1985 Oct;31(10):1720-3.
Osteocalcin, the vitamin K-dependent protein in bone, can also be detected in serum and is receiving increased attention as a marker for bone turnover in evaluating patients with metabolic bone disease. We compared results for patients as determined with two commercial radioimmunoassay kits (Immuno Nuclear Corp. and Seragen Inc.) and with our in-house radioimmunoassay (Methods Enzymol 107: 517, 1984). Results by our assay correlated well (r greater than 0.9) with those by both kits, but the values by the Immuno Nuclear and Seragen methods were respectively 40% and 10% lower than by our radioimmunoassay. Within-run variation (CV) for the two kit methods was respectively 7.3% and 9.8%, run-to-run CV was 9.7% and 8.6%. The standard curve was linear from 1 to 25 micrograms/L for the Immuno Nuclear kit, from 4 to 100 micrograms/L for the Seragen equilibrium method, and from 1 to 25 micrograms/L for the Seragen nonequilibrium method. A second freeze-thaw cycle reduced the serum values by 20% to 40% for both kit methods. A third freeze-thaw cycle further reduced values and eliminated any correlation among methods.
骨钙素是一种骨中依赖维生素K的蛋白质,在血清中也可检测到,作为评估代谢性骨病患者骨转换的标志物,它正受到越来越多的关注。我们比较了用两种商用放射免疫分析试剂盒(免疫核公司和血清基因公司)以及我们的内部放射免疫分析方法(《酶学方法》第107卷:517页,1984年)测定的患者结果。我们的分析结果与两种试剂盒的结果相关性良好(r大于0.9),但免疫核公司和血清基因公司方法测得的值分别比我们的放射免疫分析结果低40%和10%。两种试剂盒方法的批内变异系数(CV)分别为7.3%和9.8%,批间CV为9.7%和8.6%。免疫核公司试剂盒的标准曲线在1至25微克/升范围内呈线性,血清基因公司平衡法的标准曲线在4至100微克/升范围内呈线性,血清基因公司非平衡法的标准曲线在1至25微克/升范围内呈线性。两种试剂盒方法经过第二次冻融循环后,血清值降低了20%至40%。第三次冻融循环进一步降低了值,并消除了各方法之间的任何相关性。