Suppr超能文献

Genium™ 微处理器控制膝关节的改进提升了安全性,并改善了单侧和双侧使用者在假体体验方面的不同维度。

An enhancement of the Genium™ microprocessor-controlled knee improves safety and different aspects of the perceived prosthetic experience for unilateral and bilateral users.

作者信息

Klenow Tyler D, Lundstrom Russell L, Morris Arri, Patterson Stan, Simpson Chad, Trejo Ernesto G, Kannenberg Andreas

机构信息

Clinical Research & Services Department, Otto Bock HealthCare LP, Austin, TX, United States.

Clinical Services Department, Prosthetic & Orthotic Associates, Orlando, FL, United States.

出版信息

Front Rehabil Sci. 2024 May 2;5:1342370. doi: 10.3389/fresc.2024.1342370. eCollection 2024.

Abstract

INTRODUCTION

Bilateral microprocessor-controlled prosthetic knee (MPK) users have unique needs in traversing environmental barriers compared to unilateral users. An enhancement to the Genium™/Genium X3™ MPK which included an updated ruleset, hydraulics, and new bilateral parameter presets was made to improve safety while stumbling and the smoothness of gait for all users while also improving the experience of bilateral users. The purpose of the study was to evaluate the effectiveness of the enhancements in a sample with unilateral and bilateral amputation.

METHODS

A convenience sample of MPK users was recruited from two sites in the USA in two phases. Assessments included the -Test of Functional Mobility, Activity-specific Balance Confidence Scale, Prosthetic Limb User Survey of Mobility, a study-specific questionnaire, and the Comparative Activities of Daily Living (ADL) Questionnaire. Statistical significance of extracted data was tested with the Wilcoxon Rank-Sum Test for independent data and Wilcoxon Signed-Rank for paired data with an significance level of < 0.05. Unilateral subjects were age-matched to the group of bilateral subjects for between-groups and within-groups analyses.

RESULTS

Twenty-six subjects (= 26) were enrolled. Stumble frequency reduced 85% from 16.0 ± 39.7 to 2.4 ± 2.3 (= 0.008) between baseline and final assessment overall. The bilateral group reported 50% ( = 0.009) and 57% ( = 0.009) greater relative improvement in patient-reported ease and safety, respectively, of completing ADLs compared to the unilateral group. The unilateral group reported residual limb pain and low back pain reduced from 2.3 to 1.4 ( = 0.020) and 3.8 to 1.8 ( = 0.027), respectively, whereas the bilateral group did not.

DISCUSSION

Substantial reductions in stumbles, residual limb pain, and back pain were shown overall. These reductions were driven by the unilateral group who also showed improvements in comfort, exertion, and concentration while walking. The enhancements to the knee likely reduced some gait asymmetry for unilateral users. Improvements in patient-reported ease and safety of completing ADLs were shown overall and were driven by the bilateral group. This study shows further improvement in patient experience is achievable through innovation in MPK technology even for patients who appear to be functioning well.

摘要

引言

与单侧使用微处理器控制的假肢膝关节(MPK)的用户相比,双侧使用的用户在跨越环境障碍方面有独特需求。对Genium™/Genium X3™ MPK进行了改进,包括更新规则集、液压系统和新的双侧参数预设,以提高所有用户绊倒时的安全性和步态的平稳性,同时改善双侧用户的体验。本研究的目的是评估这些改进措施在单侧和双侧截肢样本中的有效性。

方法

分两个阶段从美国的两个地点招募了MPK用户的便利样本。评估包括功能活动测试、特定活动平衡信心量表、假肢使用者活动能力调查问卷、一份特定研究问卷以及比较日常生活活动(ADL)问卷。提取数据的统计学显著性通过独立数据的Wilcoxon秩和检验和配对数据的Wilcoxon符号秩检验进行,显著性水平<0.05。在组间和组内分析中,将单侧受试者与双侧受试者组进行年龄匹配。

结果

共招募了26名受试者(n = 26)。总体而言,从基线到最终评估,绊倒频率从16.0±39.7降低到2.4±2.3,降低了85%(P = 0.008)。与单侧组相比,双侧组报告在完成ADL方面患者报告的轻松程度和安全性的相对改善分别提高了50%(P = 0.009)和57%(P = 0.009)。单侧组报告残肢疼痛和腰痛分别从2.3降至1.4(P = 0.020)和3.8降至1.8(P = 0.027),而双侧组则没有。

讨论

总体上显示绊倒、残肢疼痛和背痛大幅减少。这些减少是由单侧组推动的,该组在行走时的舒适度、用力程度和注意力方面也有所改善。膝关节的改进可能减少了单侧用户的一些步态不对称。总体上显示在患者报告的完成ADL的轻松程度和安全性方面有所改善,这是由双侧组推动的。这项研究表明,即使对于功能似乎良好的患者,通过MPK技术的创新也可以进一步改善患者体验。