Europace. 2024 Jul 2;26(7). doi: 10.1093/europace/euae139.
Radiofrequency ablation is used as a first-line therapy for accessory pathways (APs). However, data regarding the effects of pulsed field ablation (PFA) on APs are limited. We sought to evaluate the acute procedural and 6-month success and safety of PFA in a cohort of patients with APs.
A focal contact force-sensing PFA catheter was used for patients with APs. Pulsed field ablation generator generated a bipolar and biphasic waveform (±1000 V) with a duration of 100 ms from the tip of the PFA catheter. A 100% acute procedural success was achieved in 10 conscious patients with APs (7 left anterolateral, 2 left inferolateral, and 1 right posteroseptal APs) including 6 (60%) patients after an initial application. The average total ablation time was 6.3 ± 4.9 s for 4.7 ± 1.8 ablation sites (ASs), including 3.1 ± 2.4 s at targets and 3.2 ± 2.9 s at 3.2 ± 2 bolus ASs. The mean skin-to-skin time was 59.3 ± 15.5 min, and PFA catheter dwell time was 29.4 ± 7.8 min. One patient encountered transient sinus arrest during PFA due to parasympathetic overexcitation. Sinus rhythm was restored in all patients without any significant adverse events during the short-term follow-up.
Pulsed field ablation of APs was feasible, effective, and safe. Its efficiency was remarkable for its ultrarapid termination of AP conduction. Further studies are warranted to prove whether utilization of PFA with current parameters can extend to manifold AP ablation.
射频消融术被用作旁路(AP)的一线治疗方法。然而,关于脉冲场消融(PFA)对 AP 影响的数据有限。我们旨在评估 PFA 在 AP 患者队列中的急性程序和 6 个月成功率及安全性。
使用聚焦接触力感应 PFA 导管治疗 AP 患者。脉冲场消融发生器从 PFA 导管尖端产生持续 100ms 的双极和双相波形(±1000V)。10 名意识清醒的 AP 患者(7 例左侧前外侧、2 例左侧下外侧和 1 例右侧后间隔 AP)均实现了 100%的急性程序成功,其中 6 例(60%)患者首次应用即成功。4.7±1.8 个消融部位(AS)的平均总消融时间为 6.3±4.9s,包括 3.1±2.4s 的靶位和 3.2±2.9s 的 3.2±2 个 bolus AS。平均皮肤-皮肤时间为 59.3±15.5min,PFA 导管驻留时间为 29.4±7.8min。1 例患者在 PFA 过程中因副交感神经过度兴奋而出现短暂性窦性停搏。所有患者在短期随访期间均无窦性节律恢复,无任何重大不良事件。
AP 的脉冲场消融是可行、有效且安全的。由于其对 AP 传导的超快速终止,其效率显著。需要进一步的研究来证明当前参数下使用 PFA 是否可以扩展到多种 AP 消融。