Suppr超能文献

评估气动涡轮洁治器和超声器械对牙根面的影响。

Evaluation of the effect on root surfaces of air turbine scalers and ultrasonic instrumentation.

作者信息

Lie T, Leknes K N

出版信息

J Periodontol. 1985 Sep;56(9):522-31. doi: 10.1902/jop.1985.56.9.522.

Abstract

Three air turbine scalers were compared to each other and to an ultrasonic instrument (CAVITRON) on medium and maximum power setting. The amount of remaining calculus, roughness and loss of tooth substance were estimated by means of well-defined index systems (RCI and RLTSI). The time required to clean the test surfaces--marked on proximal root surfaces of mandibular incisors--was also noted. Standardized scanning electron microscope (SEM) micrographs were independently "blind-scored" by three examiners, and total mean scores were calculated. The results revealed significant differences (P less than 0.001) between the instruments with respect to the amount of remaining calculus. There were also significant differences (P less than 0.05) between roughness and loss of tooth substance produced by CAVITRON at maximum power setting and that produced by the other instruments. No differences were found with regard to the time required to clean the test surface (until visually clean).

摘要

将三台空气涡轮洁治器相互比较,并与一台超声器械(CAVITRON)在中等功率和最大功率设置下进行比较。通过明确的指数系统(RCI和RLTSI)评估剩余牙石量、粗糙度和牙体组织丧失情况。还记录了清洁下颌切牙近中根面标记的测试表面所需的时间。由三名检查者对标准化扫描电子显微镜(SEM)显微照片进行独立的“盲评”,并计算总平均分。结果显示,器械之间在剩余牙石量方面存在显著差异(P<0.001)。CAVITRON在最大功率设置下产生的粗糙度和牙体组织丧失与其他器械产生的粗糙度和牙体组织丧失之间也存在显著差异(P<0.05)。在清洁测试表面所需的时间(直至肉眼可见清洁)方面未发现差异。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验