Jotikasthira N E, Lie T, Leknes K N
Department of Periodontology, School of Dentistry, University of Bergen, Norway.
J Clin Periodontol. 1992 Sep;19(8):560-9. doi: 10.1111/j.1600-051x.1992.tb00684.x.
Flat root surface areas of formalin-stored mandibular incisors with plaque and calculus were scaled by sonic (PHATELUS SONIC SCALER, SONIC FLEX 2000, TITAN-S SONIC SCALER) or ultrasonic instruments (HYGIENIST ULTRASONIC SCALER, CAVITRON) or by a new reciprocating scaling insert for the EVA/PROFIN system. The test areas were photographed by SEM and coded micrographs were independently graded by three examiners using the RCI (Remaining Calculus Index) and the RLTSI (Roughness Loss of Tooth Substance Index). The findings revealed that the sonic scalers as a group removed calculus more completely but also left significantly more roughness and loss of tooth substance than the other instruments tested. No difference was seen between the two ultrasonic scalers. The reciprocating insert gave results similar to those of the ultrasonic except for the scaling time which was significantly longer for the new "cleansing principle".
用声波器械(PHATELUS声波洁牙机、SONIC FLEX 2000、TITAN - S声波洁牙机)、超声波器械(洁牙师超声波洁牙机、CAVITRON)或用于EVA/PROFIN系统的新型往复式洁治头,对带有菌斑和牙石的福尔马林固定下颌切牙的扁平根面进行洁治。用扫描电子显微镜对测试区域拍照,三位检查者使用剩余牙石指数(RCI)和牙体组织粗糙度损失指数(RLTSI)对编码后的显微照片进行独立分级。结果显示,与其他测试器械相比,声波洁牙机作为一个组能更彻底地清除牙石,但也会留下明显更多的粗糙度和牙体组织损失。两种超声波洁牙机之间未观察到差异。除了新的“清洁原理”的洁治时间明显更长外,往复式洁治头的结果与超声波洁治头相似。