全膝关节置换术中常用的股骨组件在印度人群中适配性如何?

Does commonly used femoral components of total knee arthroplasty fit well in the Indian population?

作者信息

Bansal Shivam, Dhingra Mohit, Kalia Roop Bhushan, Saran Sonal, Barman Saptarshi, Choudhury Arghya Kundu, Regmi Anil, Niraula Bishwa Bandhu

机构信息

Department of Orthopaedics, All India Institute of Medical Sciences, Rishikesh, India.

Department of Orthopedics, AIIMS, Rishikesh, Uttarakhand, 249203, India.

出版信息

J Clin Orthop Trauma. 2024 Jun 7;53:102436. doi: 10.1016/j.jcot.2024.102436. eCollection 2024 Jun.

Abstract

BACKGROUND

20 % of patients remain dissatisfied post-Total Knee Arthroplasty. We intend to find the correlation between sizes of commonly used femoral components and morphometric data of distal femur in Indian population.

MATERIALS AND METHODS

Prospective observational study in 178 knees (178 individuals) was conducted on Magnetic Resonance Imaging scans (Axial section). The anteroposterior dimension at intercondylar notch (AP), anteroposterior dimension on the medial (MAP) and lateral (LAP) femoral condyle, total mediolateral width (ML), width of medial condyle (MLM), lateral condyle (MLL) and intercondylar distance (ICD) and aspect ratio (AR) were measured on males and females separately. These measurements were then analyzed keeping in view the available sizes of three latest knee arthroplasty systems (Attune CR, Next Gen II, Genesis II) to look for correlation and best fitting system.

RESULTS

On observation between values of male and female femur, a significant difference was seen in all the parameters (p-value<0.0001) except ICD (p-value 0.6591). On scatter plots, Attune was found to be nearer to the line of best fit compared to other two implant companies in both males and females. Outliers were highest in cases of NextGen II (45 % in males and 46 % in females) whereas it was lowest in the case of Attune (12.5 %in males and 41 % in females).

CONCLUSION

Attune CR was seen to fit best in the Indian population followed by Genesis II and NextGen II. These femoral implants are designed as per the sizes of the Western population and the Caucasians are oversized for the Indian subcontinent, especially female patients, thus needing improvement in design/sizes.

LEVEL OF EVIDENCE

III.

摘要

背景

20%的患者在全膝关节置换术后仍不满意。我们旨在找出印度人群中常用股骨组件尺寸与股骨远端形态学数据之间的相关性。

材料与方法

对178个膝关节(178名个体)进行前瞻性观察研究,采用磁共振成像扫描(轴位)。分别测量男性和女性髁间切迹前后径(AP)、股骨内侧髁前后径(MAP)、外侧髁前后径(LAP)、内外侧总宽度(ML)、内侧髁宽度(MLM)、外侧髁宽度(MLL)、髁间距离(ICD)及纵横比(AR)。然后结合三种最新膝关节置换系统(Attune CR、Next Gen II、Genesis II)的可用尺寸对这些测量值进行分析,以寻找相关性及最适配的系统。

结果

观察男性和女性股骨各项数值,除ICD(p值为0.6591)外,所有参数均存在显著差异(p值<0.0001)。在散点图上,发现Attune与其他两家植入物公司相比,在男性和女性中都更接近最佳拟合线。NextGen II的异常值最高(男性为45%,女性为46%),而Attune的异常值最低(男性为12.5%,女性为41%)。

结论

在印度人群中,Attune CR的适配性最佳,其次是Genesis II和NextGen II。这些股骨植入物是根据西方人群的尺寸设计的,对于印度次大陆的人群来说尺寸过大,尤其是女性患者,因此需要在设计/尺寸方面进行改进。

证据等级

III级。

相似文献

9
Gender differences in the knees of Chinese population.中国人群膝关节的性别差异。
Knee Surg Sports Traumatol Arthrosc. 2011 Jan;19(1):80-8. doi: 10.1007/s00167-010-1139-8. Epub 2010 Apr 21.

本文引用的文献

10
Osteoarthritis year in review 2018: rehabilitation and outcomes.2018 年骨关节炎年度回顾:康复与结局。
Osteoarthritis Cartilage. 2019 Mar;27(3):378-391. doi: 10.1016/j.joca.2018.11.010. Epub 2018 Dec 7.

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验