Phillips H O, Greene W B, Guilford W B, Mittelstaedt C A, Gaisie G, Vincent L M, Durell C
J Pediatr Orthop. 1985 Sep-Oct;5(5):546-9. doi: 10.1097/01241398-198509000-00007.
Measurement of femoral torsion by ultrasound was compared with measurements by standard roentgenographic techniques in a femoral model and in a clinical study. Magilligan, Ryder-Crane, and fluoroscopic determinations on the femoral model demonstrated acceptable interobserver and intraobserver variance. With high degrees of femoral torsion, ultrasound readings were less accurate and significantly lower than the actual angle of femoral torsion. In the clinical study, ultrasound measurements were compared with Magilligan measurements in 29 hips. Similar to the model, ultrasound readings were lower and discrepancy was greater at higher angles of torsion.
在一个股骨模型和一项临床研究中,将超声测量股骨扭转与标准X线摄影技术测量进行了比较。对股骨模型进行的马吉利根法、赖德 - 克兰法和荧光透视测定显示,观察者间和观察者内的差异均可接受。在股骨扭转角度较大时,超声读数的准确性较低,且明显低于股骨扭转的实际角度。在临床研究中,对29个髋关节的超声测量结果与马吉利根测量结果进行了比较。与模型情况类似,在扭转角度较高时,超声读数较低,差异也更大。